Frequency and time specifications – Atec Agilent-N9038A User Manual
Page 4
![background image](/manuals/318255/4/background.png)
4
Frequency range
DC coupled
AC coupled
Input 1
Option 508
Option 526
20 Hz to 8.4 GHz
20 Hz to 26.5 GHz
10 MHz to 8.4 GHz
10 MHz to 26.5 GHz
Input 2
20 Hz to 1 GHz
10 MHz to 1 GHz
Band
LO multiple (N)
0
1
20 Hz to 3.6 GHz
1
1
3.5 to 8.4 GHz
2
2
8.3 to 13.6 GHz
3
2
13.5 to 17.1 GHz
4
4
17.0 to 26.5 GHz
Frequency reference
Accuracy
± [(time since last adjustment x aging rate) + temperature stability + calibration accuracy]
Total aging
± 1 x 10
–7
/ year
± 1.5 x 10
–7
/ 2 years
Temperature stability
20 to 30 °C
Full temperature range
± 1.5 x 10
–8
± 5 x 10
–8
Achievable initial
calibration accuracy
± 4 x 10
–8
Example frequency reference
accuracy 1 year after last
adjustment
= ± (1 x 1 x 10
–7
+ 5 x
10
–8
+ 4 x 10
–8
)
= ± 1.9 x 10
–7
Residual FM
≤ (0.25 Hz x N) p-p in 20 ms (nominal)
Frequency readout accuracy (start, stop, center, marker)
± (marker frequency x frequency reference accuracy + 0.25 % x span + 5 % x RBW + 2 Hz + 0.5 x horizontal resolution
1
)
Marker frequency counter
Accuracy
± (marker frequency x frequency reference accuracy + 0.100 Hz)
Delta counter accuracy
± (delta frequency x frequency reference accuracy + 0.141 Hz)
Counter resolution
0.001 Hz
Frequency span (FFT and swept mode)
Range
0 Hz (zero span), 10 Hz to maximum frequency of instrument
Resolution
2 Hz
Accuracy
Stepped / Swept
FFT
± (0.25 % x span + horizontal resolution)
± (0.10 % x span + horizontal resolution)
1. Horizontal resolution is span/(sweep points – 1).
Frequency and Time Specifications