beautypg.com

Frequency and time specifications – Atec Agilent-N9038A User Manual

Page 4

background image

4

Frequency range

DC coupled

AC coupled

Input 1

Option 508

Option 526

20 Hz to 8.4 GHz

20 Hz to 26.5 GHz

10 MHz to 8.4 GHz

10 MHz to 26.5 GHz

Input 2

20 Hz to 1 GHz

10 MHz to 1 GHz

Band

LO multiple (N)

0

1

20 Hz to 3.6 GHz

1

1

3.5 to 8.4 GHz

2

2

8.3 to 13.6 GHz

3

2

13.5 to 17.1 GHz

4

4

17.0 to 26.5 GHz

Frequency reference

Accuracy

± [(time since last adjustment x aging rate) + temperature stability + calibration accuracy]

Total aging

± 1 x 10

–7

/ year

± 1.5 x 10

–7

/ 2 years

Temperature stability

20 to 30 °C

Full temperature range

± 1.5 x 10

–8

± 5 x 10

–8

Achievable initial

calibration accuracy

± 4 x 10

–8

Example frequency reference

accuracy 1 year after last

adjustment

= ± (1 x 1 x 10

–7

+ 5 x

10

–8

+ 4 x 10

–8

)

= ± 1.9 x 10

–7

Residual FM

≤ (0.25 Hz x N) p-p in 20 ms (nominal)

Frequency readout accuracy (start, stop, center, marker)

± (marker frequency x frequency reference accuracy + 0.25 % x span + 5 % x RBW + 2 Hz + 0.5 x horizontal resolution

1

)

Marker frequency counter

Accuracy

± (marker frequency x frequency reference accuracy + 0.100 Hz)

Delta counter accuracy

± (delta frequency x frequency reference accuracy + 0.141 Hz)

Counter resolution

0.001 Hz

Frequency span (FFT and swept mode)

Range

0 Hz (zero span), 10 Hz to maximum frequency of instrument

Resolution

2 Hz

Accuracy

Stepped / Swept

FFT

± (0.25 % x span + horizontal resolution)

± (0.10 % x span + horizontal resolution)

1. Horizontal resolution is span/(sweep points – 1).

Frequency and Time Specifications