beautypg.com

Dynamic range specifications – Atec Agilent-N9038A User Manual

Page 10

background image

10

1 dB gain compression (two-tone)

RF Input 1; RF2 Input to 1 GHz; RF Input 2 performance = RF Input 1 performance +9 dB 5 to 50 °C

Total power at input mixer
Specification

Typical

Preselector on and off,

preamp off

9 kHz to 10 MHz

10 to 500 MHz

500 MHz to 3.6 GHz

3.6 to 26.5 GHz

0 dBm

+1 dBm

0 dBm

+4 dBm (nominal)

+3 dBm (typical)

+5 dBm (typical)

+4 dBm (typical)

Preselector off, Preamp on

10 MHz to 3.6 GHz

3.6 to 26.5 GHz

Tone spacing 100 kHz to 20 MHz

Tone spacing > 70 MHz

–10 dBm (nominal)

–26 dBm (nominal)

–16 dBm (nominal)

Preselector on, Preamp on

9 kHz to 10 MHz

10 to 500 MHz

500 MHz to 3.6 GHz

3.6 to 26.5 GHz

Tone spacing 100 kHz to 20 MHz

Tone spacing > 70 MHz

–16 dBm (nominal)

–18 dBm (typical)

–16 dBm (typical)

−26 dBm (nominal)

−16 dBm (nominal)

Displayed average noise level (DANL)

(Input terminated, sample or average detector, averaging type = Log, 0 dB input attenuation, IF Gain = High, 20 to 30 °C)
RF Input 1; RF Input 2 to 1 GHz; RF Input 2 performance = RF Input 1 performance +11 dB

Specification

Typical including NFE

1

Preselector off, preamp off

20 Hz

2

100 Hz

2

1 kHz

2

9 kHz

100 kHz

1 MHz

10 MHz to 2.1 GHz

2.1 to 3.6 GHz

3.5 to 8.4 GHz

8.3 to 13.6 GHz

13.5 to 17.1 GHz

17.0 to 20.0 GHz

20.0 to 26.5 GHz

–97 dBm

–106 dBm

–118 dBm

–119 dBm

–131 dBm

–150 dBm

–150 dBm

–148 dBm

–148 dBm

–147 dBm

–141 dBm

–142 dBm

–135 dBm

–158 dBm

–157 dBm

–159 dBm

–158 dBm

–151 dBm

–152 dBm

–146 dBm

Preselector off, preamp on

100 kHz

1 MHz

10 MHz to 2.1 GHz

2.1 to 3.6 GHz

3.5 to 8.4 GHz

8.3 to 13.6 GHz

13.5 to 17.1 GHz

17.0 to 20.0 GHz

20.0 to 26.5 GHz

–144 dBm

–162 dBm

–163 dBm

–161 dBm

–164 dBm

–162 dBm

–160 dBm

–158 dBm

–155 dBm

–175 dBm

–173 dBm

–172 dBm

–173 dBm

–171 dBm

–165 dBm

–162 dBm

Dynamic Range Specifications

1. Typical DANL including NFE = Typical DANL-DANL improvement with NFE