Front panel

RF input

RF Input 1 Connector

RF Input 2 Connector

Type-N female, 50 Ω (nominal) (Standard)

3.5 mm female (Opt C35)

Type-N female, 50 Ω (nominal) (Standard)

Probe power

Voltage/current

+15 Vdc, ± 7 % at 150 mA max (nominal)

–12.6 Vdc, ± 10 % at 150 mA max (nominal)

USB 2.0 ports

Master (2 ports)

Standard

Connector

Output current

Compatible with USB 2.0

USB Type-A female

0.5 A (nominal)

Rear panel

10 MHz out

Connector

Output amplitude

Frequency

BNC female, 50 Ω (nominal)

≥ 0 dBm (nominal)

10 MHz × (1+ frequency reference accuracy)

Ext Ref In

Connector

Input amplitude range

Input frequency

Frequency lock range

BNC female, 50 Ω (nominal)

–5 to 10 dBm (nominal)

1 to 50 MHz (nominal)

± 5 x 10

–6

of specified external reference input frequency

Trigger 1 and 2 inputs

Connector

Impedance

Trigger level range

BNC female

> 10 kΩ (nominal)

–5 to 5 V

Trigger 1 and 2 outputs

Connector

Impedance

Level

BNC female

50 Ω (nominal)

0 to 5 V (CMOS)