System specifications
Scanning inputs
Analog:
34921A, 34922A, 34923A, 34924A, and
34925A multiplexer channels
Digital:
34950A/52A digital in and totalize
Scan triggering
Source
Interval, external, button press, software,
or on monitor channel alarm
Scan count
1 to 50,000 or continuous
Scan interval
0 to 99 hours; 1ms step size
Channel delay
0 to 60 seconds per channel; 1 ms step size
External trig delay
< 2 ms. With monitor on < 200 ms
External trig jitter
< 2 ms
Alarms
Analog inputs
Hi, Lo, or Hi + Lo evaluated each scan
Digital inputs
34950A/52A digital in maskable pattern
match or state change
34950A/52A frequency and totalize:
Hi limit only
Monitor channel
Alarm evaluated each reading
Alarm outputs
4 TTL compatible
Selectable TTL logic Hi or Lo on fail
Latency
5 ms (typical)
Memory
Type
Volatile
Readings
500,000 with timestamp, readable during scan
States
5 instrument states with user label
Alarm queue
Up to 20 events with channel number, reading,
and timestamp
System features
Per-channel math
Individual Mx+B scaling and calculated real
Min/max/average
time
Power fail recovery
Save switch states
Relay maintenance
Counts each relay closure and stores on
module User resettable
Real-time clock
Battery-backed, 20-year typical life
General specifications
Power supply
Universal 100 V to 240 V ± 10%
Power line frequency
50 – 60 Hz ± 10% automatically sensed
Power consumption
150 VA
Operating environment
Full accuracy for 0°C to 55°C
Full accuracy to 80% R.H. at 40°C
IEC 60664-1 pollution degree 1
Storage environment
-40°C to 70°C
[1]
Mainframe dimensions
133H x 426W x 341D mm (5.25” x 16.8” x 14”)
Full rack, 3 units high
Mainframe weight:
8.8 kg (19.6 lbs)
Module dimensions
280 x 170 x 27 mm (11” x 6.7” x 1”)
Safety conforms to
CSA, UL/IEC/EN 61010-1
EMC conforms to
IEC/EN 61326-1, CISPR 11
Warranty 1
year
Software
Agilent connectivity software included
Agilent IO Libraries Suite 14.0 (E2094N)
Minimum system requirements (IO libraries and drivers)
PC hardware
Intel Pentium 100 MHz, 64 Mbyte RAM,
210 Mbyte disk space
Display 800 x 600, 256 colors, CD-Rom drive
Operating system
[2]
Windows
®
98 SE/NT/2000/XP
Computer interfaces
Standard LAN 10BaseT/100BaseTx
Standard USB 2.0
IEEE 488.2 GPIB
Software driver support for programming languages
Software drivers:
IVI-C and IVI COM for
Windows NT/2000/XP
Compatible with programming tools and environments:
Agilent VEE Pro, Agilent T&M Toolkit
(requires Visual Studio.NET)
National Instruments Test Stand,
Measurement Studio, LabWindows/CVI,
LabVIEW, Switch Executive
Microsoft Visual Studio.NET, C/C++,
Visual Basic 6
[1] Storage at temperatures above 40°C will decrease battery life