Atec Agilent-34980A User Manual
Page 11
![background image](/manuals/318056/11/background.png)
11
Table 3. Matrix selection table—specifications and characteristics*
34931A
34932A
34933A
Channels/configurations
dual 4x8
dual 4x16
dual 4x8
8x8
8x16
8x8
4x16
4x32
4x16
quad 4x8, 1-wire
Switch type
Armature
Armature
Reed
latching
latching
non-latching
Input characteristics (per channel)
Max volts
± 300 V
[1]
± 300 V
[1]
± 150 V peak
[2]
Max current (DC, AC RMS)
Switch current
1 A
1 A
0.5 A
Carry current
2 A
2 A
1.5 A
Power (W, VA)
[2]
60 W
60 W
10 W
[7]
Volt-Hertz limit
10
8
10
8
10
8
General Specifications
Offset voltage
[3]
< 3 uV
< 3 uV
< 50 uV
< 100 uV 1-wire
Initial closed channel res
[3]
< 1.5
Ω
< 1.5
Ω
< 1.5
Ω
[5]
AC characteristics
Bandwidth at terminal block
[4]
30 MHz
30 MHz
30 MHz
[5]
2 MHz 1-wire
Crosstalk at terminal block (ch-ch)
[4]
300 kHz
-65 dB
-65 dB
-65 dB
1 MHz
-55 dB
-55 dB
-55 dB
20 MHz
-30 dB
-30 dB
-40 dB
Capacitance at terminal block
HI-LO
50 pF
50 pF
80 pF
LO – earth
80 pF
80 pF
75 pF
General characteristics
Relay life, typical
No load
100 M
100 M
1000 M
10 V, 100 ma
10 M
10 M
10 M
Rated load
100 k
100 k
10 k
Open/close time
4 ms/4 ms
4 ms/4 ms
0.5 ms/0.5 ms
Analog bus backplane connection
Bank 2
Bank 2
Bank 2
[1] DC or AC RMS voltage, channel-to-channel or channel-to-earth
[2] Peak voltage, channel-to-channel or channel-to-earth
[3] Into analog bus. System errors are included in the internal DMM
measurement accuracy specifications
[4] 50
Ω source, 50 Ω load, differential measurements verified (Sdd21)
[5] With in-rush resistors bypassed
[6] Limited to 6 W channel resistance power loss per module
[7] Power restrictions allow only 20 channels to be closed at one time
* See User’s Guide for additional specifications