Atec Agilent-34980A User Manual

Page 20

background image

20

34980A system specifications and characteristics

(for complete specifications, see the 34980A Users Guide)

DMM accuracy ± (% of reading + % of range)
Includes measurement error, switching error, and transducer conversion error

Measurement including switch error

[1]

Function

Range

[4]

Frequency, etc.

24 hour

[2,3]

90 days

1 year

Temperature

Tcal ± 1°C

Tcal ± 5°C

Tcal ± 1°C

coefficient
Tcal ± 5°C

DC voltage

100.0000 mV

0.0030 + 0.0035

0.0040 + 0.0040

0.0050 + 0.0040

0.0005 + 0.0005

(with 34921A/22A/

1.000000 V

0.0020 + 0.0006

0.0030 + 0.0007

0.0040 + 0.0007

0.0005 + 0.0001

31A/32A)

10.00000 V

0.0015 + 0.0004

0.0020 + 0.0005

0.0035 + 0.0005

0.0005 + 0.0001

100.0000 V

0.0020 + 0.0006

0.0035 + 0.0006

0.0045 + 0.0006

0.0005 + 0.0001

300.0000 V

0.0020 + 0.0020

0.0035 + 0.0030

0.0045 + 0.0030

0.0005 + 0.0003

True RMS AC

100.0000 mV

3Hz-5Hz

1.00 + 0.03

1.00 + 0.04

1.00 + 0.04

0.100 + 0.004

voltage

[5]

to 100.0000 V

5Hz-10Hz

0.35 + 0.03

0.35 + 0.04

0.35 + 0.04

0.035 + 0.004

10Hz-20kHz

0.04 + 0.03

0.05 + 0.04

0.06 + 0.04

0.005 + 0.004

20kHz-50kHz

0.10 + 0.05

0.11 + 0.05

0.12 + 0.05

0.011 + 0.005

50kHz-100kHz

0.55 + 0.08

0.60 + 0.08

0.60 + 0.08

0.060 + 0.008

100kHz-300kHz

[6]

4.00 + 0.50

4.00 + 0.50

4.00 + 0.50

0.20 + 0.02

300.0000 V

3Hz-5Hz

1.00 + 0.05

1.00 + 0.08

1.00 + 0.08

0.100 + 0.008

5Hz-10Hz

0.35 + 0.05

0.35 + 0.08

0.35 + 0.08

0.035 + 0.008

10Hz-20kHz

0.04 + 0.05

0.05 + 0.08

0.06 + 0.08

0.005 + 0.008

20kHz-50kHz

0.10 + 0.10

0.11 + 0.12

0.12 + 0.12

0.011 + 0.012

50kHz-100kHz

0.55 + 0.20

0.60 + 0.20

0.60 + 0.20

0.060 + 0.020

100kHz-300kHz

[6]

4.00 + 1.25

4.00 + 1.25

4.00 + 1.25

0.20 + 0.05

Resistance

[7]

100.0000

1mA

0.0030 + 0.0035

0.008 + 0.004

0.010 + 0.004

0.0006 + 0.0005

1.000000 k

1mA

0.0020 + 0.0006

0.008 + 0.001

0.010 + 0.001

0.0006 + 0.0001

10.00000 k

100uA

0.0020 + 0.0005

0.008 + 0.001

0.010 + 0.001

0.0006 + 0.0001

100.0000 k

10uA

0.0020 + 0.0005

0.008 + 0.001

0.010 + 0.001

0.0006 + 0.0001

1.000000 M

5.0uA

0.002 + 0.001

0.008 + 0.001

0.010 + 0.001

0.0010 + 0.0002

10.00000 M

500nA

0.015 + 0.001

0.020 + 0.001

0.040 + 0.001

0.0030 + 0.0004

100.0000 M

500nA/10M

0.300 + 0.010

0.800 + 0.010

0.800 + 0.010

0.1500 + 0.0002

Frequency and

100mV to 300V

3Hz-5Hz

0.10

0.10

0.10

0.005

period

[8]

5Hz-10Hz

0.05

0.05

0.05

0.005

10Hz-40Hz

0.03

0.03

0.03

0.001

40Hz-300kHz

0.006

0.01

0.01

0.001

DC current

10.00000 mA

< 0.1 V burden

0.005 + 0.010

0.030 + 0.020

0.050 + 0.020

0.002 + 0.0020

(34921 only)

100.0000 mA

< 0.6V

0.010 + 0.004

0.030 + 0.005

0.050 + 0.005

0.002 + 0.0005

1.000000 A

< 2V

0.050 + 0.006

0.080 + 0.010

0.100 + 0.010

0.005 + 0.0010

True RMS AC current

10.00000 mA

3Hz-5Hz

1.00 + 0.04

1.00 + 0.04

1.00 + 0.04

0.100 + 0.006

(34921A only)

and

[5]

5Hz-10Hz

0.30 + 0.04

0.30 + 0.04

0.30 + 0.04

0.035 + 0.006

1.0 A

10Hz-5kHz

0.10 + 0.04

0.10 + 0.04

0.10 + 0.04

0.015 + 0.006

100.0000 mA

[9]

3Hz-5Hz

1.00 + 0.5

1.00 + 0.5

1.00 + 0.5

0.100 + 0.006

5Hz-10Hz

0.30 + 0.5

0.30 + 0.5

0.30 + 0.5

0.035 + 0.006

10Hz-5kHz

0.10 + 0.5

0.10 + 0.5

0.10 + 0.5

0.015 + 0.006

[1] One hour warm-up and a fixed configuration

with slow AC filter, sine wave input, and 6

1

2

digits. Temperature within ± 5°C of tempera-
ture at calibration (Tcal between 18-28°C).

[2] 90 minute warm-up and a fixed configuration

and 6

1

2

digits. Temperature within ± 1°C of

temperature at calibration (Tcal between
18-28°C).

[3] Relative to calibration standards

[4] 20% over range on all ranges except 300VDC

and AC ranges and 1 ADC and AC current
ranges

[5] For singe wave input > 5% of range. For

inputs from 1% to 5% of range and < 50 kHz
add 0.1% of range additional error.
For AC filter slow.

[6] Typically 30% of reading error at 1 MHz,

limited to 1 x 10

8

volt-hertz

[7] For 4-wire ohms or 2-wire ohms using scaling

to remove offset. Add 4 ohms additional error
to 2-wire ohms function without scaling.

[8] Input > 100 mV. For 10 mV inputs multiply

% of reading error x 10. For 1 sec aperture.

[9] Specified only for inputs > 10 mA.

For AC filter slow.