Automatic limit testing of
Maximum Tolerable Wander, MTW
ITU-T G.823, G.824
The ANT-20SE tests the device under test for conformance to the
standard tolerance mask limits for maximum tolerable wander.
Measurement poinits . . . . . . . . . up to 10 Frequency/Amplitude values
Detection criteria . . . . . . . . . . . . . . . . . . . . . . . . . TSE (bit error), alarms
Frequency range . . . . . . . . . . . . . . . . . . . . . 10 mHz to 10 Hz, step 1 mHz
Amplitude range . . . . . . . . . . . . . . . . . . 0.1 to 200 000 UI, step: 0.1 UI
The result of each measurement is shown in a table with an ªOKº or
ªFAILEDº message.
Accessory
Acterna TSR-37
DA 3700/00
Rubidium Timing Signal Reference
The TSR-37 is a powerful reference source to quickly measure and test
the synchronization quality of PDH/SDH/SONET digital networks.
MTIE and TDEV measurements for up to 1000 seconds can be easily
performed without a GPS reference. Coupled with the optional
GPS-FC, the range of observation time can be largely extended to meet
specific requirements.
Provides the reference clock for wander analysis using the ANT-20.
. PDH/SDH/SONET Wander measurement source
. Accuracy at 25 C: +5610
±11
without GPS
51610
±11
with GPS
. 12 outputs, framed and unframed:
5 MHz, 10 MHz, 2.048 kHz, 1.544 kHz, E1, T1
. Compact, robust & lightweight
. External autocalibration input
. Input for GPS or Cesium reference
See Acterna TSR-37 data sheet for details.