Specifications, Mainframe specifications, General specifications – Atec Agilent-83491A-92A User Manual
Page 8
Specifications
Mainframe Specifications
HORIZONTAL SYSTEM (Time Base)
Scale Factor (Full scale is ten divisions.)
Minimum
2 ps/div (with 86107A: 500 fs/div)
Maximum
1 s/div
Delay
1
Minimum
≥24 ns (Standard), 28 ns (Option 001)
Maximum
1000 screen diameters or 10 s, whichever is smaller
Time Interval Accuracy
≤8 ps + 0.1% of reading (dual marker measurement performed at a temperature within
±5°C of horizontal calibration temperature)
8 ps + 0.1% + 0.5 ps/°C to (5 < absolute temperature value <15°C)
Time Interval Resolution
≤ (screen diameter)/(record length) or 62.5 fs, whichever is larger
Display Units
Bits or time (TDR mode–meters)
VERTICAL SYSTEM (Channels)
Number of Channels
4 (simultaneous acquisition)
Vertical Resolution
12 bit A/D converter (up to 15 bits with averaging)
Full Resolution Channel Scales
Adjusts in a 1-2-5-10 sequence for coarse adjustment or fine adjustment resolution
from the front panel knob
Adjustments
Scale, offset, activate filter, sampler bandwidth, attenuation factor, transducer conversion factors
Typical Acquisition Rate
40 kHz per Channel
Record Length
Manual setting between 450 and 4050 samples (increments of 1, 1350 default sample setting)
or using up arrow/down arrow keys, increments of 450 samples: 450, 900, 1350 (default) etc.
up to 4050 samples
8
Specifications describe warranted performance over the temperature range of +10°C to + 40°C (unless otherwise noted). The specifications are
applicable for the temperature after the instrument is turned on for one (1) hour, and while self-calibration is valid. Many performance parameters are
enhanced through frequent, simple user calibrations. Characteristics provide useful, non-warranted information about the functions and
performance of the instrument. Characteristics are printed in italic typeface.
Factory Calibration Cycle -For optimum performance, the instrument should have a complete verification of specifications once every twelve (12) months.
General Specifications
This instrument meets Agilent Technologies’ environmental specifications (section 750) for class B-1 products with exception as described for
temperature and condensation. Contact your local field engineer for complete details.
Temperature
Operating
10°C to +40°C (50°F to + 104°F)
Non-operating
–40°C to +70°C (–40°F to +158°F)
Humidity
Operating
Up to 90% humidity (non-condensing) at +40°C (+104°F)
Non-operating
Up to 95% relative humidity at +65°C (+149°F)
Altitude
Operating
Up to 4,600 meters (15,000 ft)
Non-operating
Up to 15,300 meters (50,000 ft)
Vibration
Operating
Random vibration 5–500 Hz, 10 minutes per axis, 2.41 g (rms)
Non-operating
Random vibration 5–500 Hz, 10 minutes per axis, 0.3 g (rms); Resonant search, 5–500 Hz
swept sine, 1 octave/min sweep rate, 0.75 g, 5 minute resonant dwell at 4 resonances/raxis
Power Requirements
Voltage
90–132 or 198–264 Vac, 48–66 Hz
Power (including modules)
604 VA; 391 W
Weight
Mainframe without modules
15.5 kg (34 lb)
Each Module
1.2 kg (2.6 lb)
Mainframe Dimensions (excluding handle)
Without front connectors and rear feet
215.1 mm H x 425.5 mm W x 566 mm D (8.47 in x 16.75 in x 22.2 in)
With front connectors and rear feet
215.1 mm H x 425.5 mm W x 629 mm D (8.47 in x 16.75 in x 24.8 in)
1 Time offset relative to the front panel trigger output on the instrument mainframe.