High Bandwidth, Single-Mode
Optical/Electrical Modules
86106B
86109A 86109B
86116A
1
86116B
1
OPTICAL CHANNEL SPECIFICATIONS
Optical Channel Unfiltered Bandwidth
28 GHz
30 GHz
40 GHz
2
53 GHz
65 GHz (best pulse fidelity)
Wavelength Range
1000–1600 nm
55 GHz (best sensitivity)
Calibrated Wavelengths
1310/1550 nm
1480-1620 nm
Optical Sensitivity
4
–7 dBm
N/A
Transition Time (10% to 90%,
calculated from TR = 0.48/BW optical)
18 ps
16 ps
12 ps (FWHM)
3
9.0 ps (FWHM)
3
7.4 ps (FWHM)
RMS Noise
Characteristic
13 µW (Filtered)
12 µW
25 µW (30 GHz)
60 µW (50 GHz)
140 µW (65 GHz)
23 µW (Unfiltered)
65 µW (40 GHz)
190 µW (53 GHz)
50 µW (55 GHz)
Maximum
15 µW (Filtered)
30 µW
30 µW (30 GHz)
90 µW (50 GHz)
250 µW (65 GHz)
30 µW (Unfiltered)
75 µW (40 GHz)
260 c(53 GHz)
85 µW (55 GHz)
Scale Factor
Minimum
20 µW/division
200 µW/division
Maximum
500 µW/division
1.0 mW/division
2.5 mW/division
5 mW/division
CW Accuracy (single marker,
±50 µW ±4% of
referenced to average power monitor)
(reading-channel offset)
±150 µW ±4% of (reading-channel offset)
CW Offset Range (referenced two
divisions from screen bottom)
+1 mW to –3 mW
+6 mW to –2 mW
+5 mW to –15mW
+8 to –12 mW
Average Power Monitor
(specified operating range)
–27 dBm to +3 dBm
–23 dBm to + 9 dBm
Factory Calibrated Accuracy
±5% ±100 nW ±connector uncertainty, 20°C to 30°C
User Calibrated Accuracy
±2% ±100 nW ±power meter uncertainty, <5°C change
Maximum Input Power
Maximum non-destruct average
2 mW (+3 dBm)
10 mW (+10 dBm)
Maximum non-destruct peak
10 mW (+10 dBm)
50 mW (+17 dBm)
Fiber Input
9/125 µm, user selectable connector
Input Return Loss
(HMS-10 connector fully filled fiber)
30 dB
20 dB
1
86116A requires the 86100A/B software revision A.3.0 or above.
2
Specified with 8 point moving average in frequency response.
3
FWHM (Full Width Half Max) as measured from optical pulse with 700 fs FWHM, 5 MHz repetition rate and 10 mW peak power.
4
Smallest average optical power required for mask test. Valves represent typical sensitivity of NRZ eye diagrams. Assumes mask test with compliance filter switched in.
ELECTRICAL CHANNEL SPECIFICATIONS
Electrical Channel Bandwidth
18 and 40 GHz
26 and 50 GHz
43 and 63 GHz
80, 55 and 30 GHz
Transition Time (10% to 90%,
19.5 ps (18 GHz)
<13.2 ps (26 GHz)
8.1 ps (43 GHz)
6.4 ps (55 GHz)
calculated from TR = 0.35/BW)
9 ps (40 GHz)
7 ps (50 GHz)
5.6 ps (63 GHz)
4.4 ps (80 GHz)
RMS Noise
Characteristic
0.25 mV (18 GHz)
0.25 mV (26 GHz)
0.6 mV (43 GHz)
0.6 mV (55 GHz)
0.5 mV (40 GHz)
0.60 mV (50 GHz)
1.7 mV (63 GHz)
1.1 mV (80 GHz)
Maximum
0.5m V (18 GHz)
0.50 mV (26 GHz)
0.9 mV (43 GHz)
1.1 mV (55 GHz)
1.0 mV (40 GHz)
1.0 mV (50 GHz)
2.5 mV (63 GHz)
2.2 mV (80 GHz)
Scale Factor
Minimum
1 mV/division
2 mV/division
Maximum
100 mV/division
100 mV/division
DC Accuracy (single marker)
±0.4% of full scale
±0.4% of full scale
±0.8% of full scale
±0.4% of full scale
±2 mV ±1.5% of (reading-
±2 mV ±1.5% of (reading- ±2 mV ±1.5% of (reading- ±3 mV ±2% of (reading-
channel offset), 18 GHz
channel offset), 26 GHz
channel offset), 43 GHz
channel offset), ±2% of
±0.4% of full scale
±0.4% of full scale
±2.5% of full scale
offset (all bandwidths)
±2 mV ±3% of (reading-
±2 mV ±2% of (reading- ±2 mV ±2% of (reading-
channel offset), 40 GHz
channel offset), 50 GHz
channel offset), 63 GHz
DC Offset Range (referenced
to center of screen)
±500 mV
Input Dynamic Range
(relative to channel offset)
±400 mV
Maximum Input Signal
±2 V (+16 dBm)
Nominal Impedance
50 ohm
Reflections (for 20 ps rise time)
5%
10% (DC–70 GHz)
20% (70–100 GHz)
Electrical Input
2.4 mm (male)
1.85 mm (male)