beautypg.com

Technical bulletin, Detect mode troubleshooting – Franklin Fueling Systems TS-LS300 Troubleshooting Tips for the LS300 Line Leak Detection System User Manual

Page 2

background image

TECHNICAL BULLETIN

TB0505-01

May 09, 2005

Franklin Fueling Systems 3760 Marsh Rd. Madison WI 53718 USA

Tel: +1 608 838 8786 800 225 9787 Fax: +1 608 838 6433 www.franklinfueling.com

Detect Mode Troubleshooting

Condition

Definition

Potential Causes

Proposed Action

Leak Alarm

• Gross (3gph) Leak

Alarm


The LS300 has detected
a pressure loss
consistent with a leak of
3gph or greater.

• Monthly or Annual

Precision Test Failed

The LS300 has detected
a pressure loss
consistent with a .2gph
leak (one flash) or a
.1gph leak (two flashes).

• Leak in the pipeline

• Submersible pump check

valve not seating properly

• Leak in the manual

pressure relief (FE Petro
Pump)

• Adjustable Functional

Element not seating
properly/defective. (Red
Jacket)

• Push the Test/Reset button to clear alarm.

Conduct a manual leak test to confirm the line
status. If the LS300 again detects a leak after
three failed test sessions, shut down the line and
troubleshoot the pressure loss.

• Check valve seating area should be clean and free

of debris. Clean area if necessary.

• Verify that the manual pressure relief screw (FE

Petro) is turned clockwise.

• Check sump areas for signs of leakage.

• For further information, see FE Petro Service

Bulletin SB007 STP/IST Static Pressure Testing.

Transducer Not Seen

The LS300 cannot
detect the signal coming
from the transducer

• Transducer is wired

incorrectly

• Wiring or connections to the

transducer are loose or
broken

• Transducer is defective

• See the LS300 Auto-Learn User’s Guide‘s

Transducer Troubleshooting section for specific
troubleshooting information.

Pipeline Failed to Hold
Pressure

After the submersible
pump shut off, pipeline
pressure dropped
immediately to 0 psi.

• Large leak in the pipeline

• Check valve removed or

stuck open

• Press Test/Reset button to clear alarm. Conduct a

manual 3gph test and confirm line status.

• Verify that the check valve is installed in the

submersible pump and is seating properly.

• Check sump areas for signs of leakage.

• For further information, see FE Petro Service

Bulletin SB007 STP/IST Static Pressure Testing.

Insufficient Pressure to
Conduct Test

Pipeline is not fully
pressurizing when the
submersible pump is
turned on.

• Power to submersible is

OFF

• Leak in the pipeline

• Transducer problem

• Verify that there is power to the submersible

pump.

• Check sump areas for signs of a large leak.

• See the LS300 Auto-Learn User’s Guide‘s

Transducer Troubleshooting section for specific
troubleshooting information.

Calibration is Incomplete

The line will not detect
leaks because the line is
not Learned

• Learn sequence was not

attempted or was not
completed properly

• See the LS300 Auto-Learn User’s Guide‘s Line

Calibration. Follow the instructions on learning a
line.


Highly flammable vapors or liquids may be present in the environment in which this equipment is
installed or serviced. Installing or working on this equipment means working in an environment that
presents risks of severe injury or death if instructions and standard industry practices are not followed.
Follow all applicable codes governing the installation and servicing of this product and the entire system.
Always lock out and tag electrical circuit breakers while installing or servicing this equipment and related
equipment. Refer to the Installation and Owner’s Manual
of this equipment and related equipment for
complete installation and safety information.

Warning