What to do if...

ba77064e01

03/2014

59

11 What to do if...

11.1 pH

Error message

OFL, UFL

The measured value is outside the measuring range.

Error message,

Error

More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.

Cause

Remedy

IDS pH sensor:

– The measured value is outside

the measuring range of the
meter

– Use suitable IDS pH sensor

– Air bubble in front of the junction – Remove air bubble

(e.g. sway or stir the solution)

– Gel electrolyte dried out

– Replace the sensor

Cause

Remedy

IDS pH sensor:

– The values determined for zero

point and slope of the IDS pH
sensor are outside the allowed
limits.

– Recalibrate

– Sensor broken

– Replace the sensor

Buffer solutions:

– The used buffer solutions do not

agree with the set buffer set

– Set different buffer set
or
– Use different buffer solutions

– Buffer solutions too old

– Use only once.

Note the shelf life

– Buffer solutions depleted

– Change solutions