Xylem HandyLab 600 User Manual
Page 28

pH value
HandyLab 600
28
ba77064e01
03/2014
Continuing with two-
point calibration
7.
Wait for the measurement with stability control to be completed or ter-
minate the stability control and take over the calibration value with
The pH value of the buffer solution is displayed.
8.
Set the nominal buffer value for the measured temperature with
<><>.
9.
Accept the calibration value with
The calibration display for the next buffer appears (voltage display).
10. If necessary, finish the calibration procedure as a single-point calibration
with
The calibration record is displayed.
For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the IDS
pH sensor.
pH
0.0
24.8 °C
ConCal
7.000
01.07.2011
08:00
Buffer
mV
AR
pH
7.000
24.8 °C
ConCal
1
01.07.2011
08:00
Buffer
11. Thoroughly rinse the sensor with deionized water.
12. Immerse the pH sensor in buffer solution 2.
13. Start the measurement with
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.