4 reset, 1 resetting the measurement settings, Reset – Xylem HandyLab 600 User Manual
Page 47: Resetting the measurement settings, Sec- tion 7.4.1 r, Esetting, Measurement, Settings, Section 7.4.1 r

HandyLab 600
Settings
ba77064e01
03/2014
47
7.4
Reset
You can reset (initialize) all sensor settings and sensor-independent settings
separately from each other.
7.4.1 Resetting the measurement settings
pH
The following settings for pH measurements are reset to the default settings
with the Reset function:
The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the
key.
ORP
The following settings for ORP measurements are reset to the default settings
with the Reset function:
The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the
key.
The calibration data are reset to the default settings together with
the measuring parameters. Recalibrate after performing a reset.
Setting
Default settings
Buffer
Calibration interval
7 d
Unit for slope
mV/pH
Measured parameter
pH
Unit for slope
0.001
mV resolution
0.1
Asymmetry
0 mV
Slope
-59.2 mV
Man. temperature
25 °C
Single-point calibration
off
Setting
Default settings
mV resolution
0.1
Man. temperature
25 °C