5 calibration points, 6 calibration data, Calibration points – Xylem HandyLab 600 User Manual
Page 30: Calibration data

pH value
HandyLab 600
30
ba77064e01
03/2014
5.2.5 Calibration points
Calibration can be performed using one to five buffer solutions in any order
(single-point to five-point calibration). The meter determines the following
values and calculates the calibration line as follows:
5.2.6 Calibration data
22. Set the nominal buffer value for the measured temperature with
<><>.
23. Accept the calibration value with
The calibration display for the next buffer appears (voltage display).
24. If necessary, use
The calibration record is displayed.
or
Continue calibrating using the next buffer with
After the fifth buffer has been measured the calibration is automat-
ically finished. Then the calibration record is displayed.
The calibration line is determined by linear regression.
Calibration
Determined
values
Displayed calibration data
1-point
Asy
Zero point = Asy
Slope = Nernst slope
(-59.2 mV/pH at 25 °C)
2-point
Asy
Slp.
Zero point = Asy
Slope = Slp.
3-point to 5-
point
Asy
Slp.
Zero point = Asy
Slope = Slp.
The calibration line is calculated by
linear regression.
You can display the slope in the units, mV/pH or % (see section
7.1.1 S
The calibration record is automatically transmitted to the interface
after calibrating.