5 calibration points, 6 calibration data, Calibration points – Xylem HandyLab 600 User Manual

Page 30: Calibration data

background image

pH value

HandyLab 600

30

ba77064e01

03/2014

5.2.5 Calibration points

Calibration can be performed using one to five buffer solutions in any order
(single-point to five-point calibration). The meter determines the following
values and calculates the calibration line as follows:

5.2.6 Calibration data

22. Set the nominal buffer value for the measured temperature with

<><>.

23. Accept the calibration value with .

The calibration display for the next buffer appears (voltage display).

24. If necessary, use to finish the calibration.

The calibration record is displayed.
or
Continue calibrating using the next buffer with .

After the fifth buffer has been measured the calibration is automat-
ically finished. Then the calibration record is displayed.

The calibration line is determined by linear regression.

Calibration

Determined
values

Displayed calibration data

1-point

Asy

Zero point = Asy

Slope = Nernst slope
(-59.2 mV/pH at 25 °C)

2-point

Asy
Slp.

Zero point = Asy

Slope = Slp.

3-point to 5-
point

Asy
Slp.

Zero point = Asy

Slope = Slp.

The calibration line is calculated by
linear regression.

You can display the slope in the units, mV/pH or % (see section
7.1.1 S

ETTINGS

FOR

P

H

MEASUREMENTS

, page 40).

The calibration record is automatically transmitted to the interface
after calibrating.