Xylem HandyLab 600 User Manual
Page 29

HandyLab 600
pH value
ba77064e01
03/2014
29
Continuing with
three- to five-point
calibration
14. Wait for the measurement with stability control to be completed or ter-
minate the stability control and take over the calibration value with
The pH value of the buffer solution is displayed.
15. Set the nominal buffer value for the measured temperature with
<><>.
16. Accept the calibration value with
The calibration display for the next buffer appears (voltage display).
17. If necessary, finish the calibration procedure as a two-point calibration
with
The calibration record is displayed.
pH
4.035
24.8 °C
ConCal
2
01.07.2011
08:00
Buffer
18. Thoroughly rinse the sensor with deionized water.
19. Immerse the sensor in the next buffer solution.
20. Start the measurement with
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
21. Wait for the measurement with stability control to be completed or ter-
minate the stability control and take over the calibration value with
The pH value of the buffer solution is displayed.
pH
9.958
24.8 °C
ConCal
3
01.07.2011
08:00
Buffer