Xylem HandyLab 600 User Manual
Page 26

pH value
HandyLab 600
26
ba77064e01
03/2014
Continuing with
three- to five-point
calibration
12. Start the measurement with
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
13. Wait for the measurement with stability control to be completed or ter-
minate the stability control and take over the calibration value with
The calibration display for the next buffer appears (voltage display).
14. If necessary, finish the calibration procedure as a two-point calibration
with
The calibration record is displayed.
pH
0.0
24.8 °C
7.000
01.07.2011
08:00
AR
Buffer
mV
15. Thoroughly rinse the sensor with deionized water.
16. Immerse the sensor in the next buffer solution.
17. Start the measurement with
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
18. Wait for the measurement with stability control to be completed or ter-
minate the stability control and take over the calibration value with
The calibration display for the next buffer appears (voltage display).
pH
180.0
24.8 °C
01.07.2011
08:00
AR
Buffer
mV