Ph 1 – Xylem HandyLab 600 User Manual
Page 25

HandyLab 600
pH value
ba77064e01
03/2014
25
Continuing with two-
point calibration
4.
Thoroughly rinse the sensor with deionized water.
5.
Immerse the sensor in the first buffer solution.
6.
For measurements without temperature sensor :
Enter the temperature of the buffer with <><>.
7.
Start the measurement with
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
8.
Wait for the end of the measurement with stability control or accept the
calibration value with
The calibration display for the next buffer appears (voltage display).
9.
If necessary, finish the calibration procedure as a single-point calibra-
tion with
The calibration record is displayed.
For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the IDS
pH sensor.
pH
1
-180.0
24.8 °C
01.07.2011
08:00
Buffer
mV
pH
-180.0
24.8 °C
01.07.2011
08:00
AR
Buffer
mV
10. Thoroughly rinse the sensor with deionized water.
11. Immerse the pH sensor in buffer solution 2.