6 orp, 1 measuring, 1 measuring the orp – Xylem HandyLab 600 User Manual
Page 37: Measuring, Measuring the orp, 6orp
HandyLab 600
ORP
ba77064e01
03/2014
37
6
ORP
6.1
Measuring
6.1.1 Measuring the ORP
NOTE
When connecting a grounded PC/printer, measurements cannot be
performed in grounded media as the values would be incorrect. The USB
interface is not galvanically isolated.
Stability control
(AutoRead)
& HOLD function
The stability control function (AutoRead) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproduc-
ibility of measured values.
The measured parameter flashes on the display
as soon as the measured value is outside the stability range
when the automatic Stability control is switched off.
You can start the Stability control manually at any time, irrespective of the set-
ting for automatic Stability control (see section 7.3.3 A
, page 46) in the System menu.
IDS ORP sensors are not calibrated. However, you can check IDS
ORP sensors using a test solution.
1.
Connect the ORP sensor to the meter.
The ORP measuring window is displayed.
2.
If necessary, check the ORP sensor using a check solution.
3.
Immerse the ORP sensor in the test sample.
U
157.0
mV
24.8 °C
01.02.2014
08:00
Info
USB output
1.
Freeze the measured value with
The [HOLD] status indicator is displayed.
The HOLD function is active.