Powersolve PPT30ME Series User Manual
Medical plug top power adaptors, Features
Powersolve Electronics Ltd. 8A Arnhem Road, Newbury, RG14 5RU. England
Tel: 01635 521858 Fax: 01635 523771 Email: [email protected]
Medical Plug Top Power Adaptors
PPT30ME Series
30 Watt Switch Mode
Universal AC Input
INPUT
Input Voltage Range
90 - 264Vac
Frequency
47 - 63Hz
Input Current
0.4 - 0.8A max.
Inrush Current
50A max. @ 240Vac
Conducted EMI
CISPR/FCC Class B
Isolation - Input/Output
5,656VDC
Leakage Current
100µA max.
OUTPUT
Hold-up Time
10mS typ. @ 115Vac
Short Circuit Protection
Continuous (Auto Recovery)
Over Voltage Protection
Yes
ENVIRONMENTAL
Operating Temperature Range
0°C to +40°C
Storage Temperature Range
-20°C to +85°C
Cooling
Natural convection
GENERAL
Safety Approvals
cUL/UL 60950, TUV EN60950, cUL/UL2601-1, TUV EN60601-1
Dimensions
108.67 x 61.98 x 36.70 mm
4.270 x 2.440 x 1.445 inches
Weight
300g
Electrical Specification
The PPT30ME series of CEC level IV compliant AC-DC
Power Adaptors offer user interchangeable AC input
plugs to enable operation in many different countries.
Adaptors can be ordered with up to 4 different
interchangeable plugs to meet the power socket
requirements of most countries.
Features
•
Universal Input Range 90-264VAC
•
Over Voltage & Continuous Short Circuit Protection
•
Meets EN60601-1 & EN55011 Class B
•
CEC Level IV Compliant (Output Cable Length ≤1800mm)
(PPT30ME-050: Output Cable Length ≤1500mm)
•
Optional Output Connector
www.powersolve.co.uk
Output Voltage and Current Ratings
MODEL
OUTPUT
OUTPUT CURRENT
RIPPLE &
VOLTAGE
LINE
LOAD
% EFF.
VOLTAGE
Min.
Max.
NOISE
ACCURACY
REG.
REG.
PPT30ME-050
5V
0A
4.0A
1%
±2%
±1%
±6%
73%
PPT30ME-090
9V
0A
3.0A
1%
±2%
±1%
±3%
76%
PPT30ME-120
12V
0A
2.5A
1%
±2%
±1%
±2%
79%
PPT30ME-150
15V
0A
2.0A
1%
±2%
±1%
±2%
81%
PPT30ME-180
18V
0A
1.67A
1%
±2%
±1%
±2%
81%
PPT30ME-240
24V
0A
1.25A
1%
±2%
±1%
±2%
83%
NOTE:
1. Voltage set point at 60% full load.
2. Output noise and ripple measured with 10µF electrolytic & 0.1µF ceramic capacitors across output at 20MHz BW.
3. Line regulation is measured from 100Vac to 240Vac at full load.
3. Load regulation is measured from 60% to full load and from 60% to 20% load (60% ±40% load).
Typical at 25°C, nominal line and 75% load unless otherwise stated.