Powersolve Electronics Ltd. 8A Arnhem Road, Newbury, RG14 5RU. England

Tel: 01635 521858 Fax: 01635 523771 Email: [email protected]

www.powersolve.co.uk

AC-DC Converter

Input Voltage

90-264VAC full range, 47-63Hz / 127-370VDC. Below 100VAC, derate load linearly to 90% at 90VAC.

Power Factor

0.95 at 230VAC / 0.99 at 115VAC at full load

AC Input Current (typ.)

14.5A at 115VAC / 6A at 230VAC

Inrush Current

30A at 115VAC / 45A at 230VAC

Leakage Current

<300µA at 264VAC for Earth / <100µA at 264VAC for Patient

Output Voltage

See Table

Output Current

See Table

Voltage Tolerance

±2%

Voltage Adjust Range

±5.0% typical adjustment by potentiometer

Overload Protection

>105% of rated output power. Protection type: Constant current limiting

Overvoltage Protection

Variable OVP, 125% ±7% of Vout. Output latches off, recycle AC input to recover or inhibit

Over Temperature Protection

Shuts down output, recovers automatically (100°C ±5°C detect on heatsink on secondary side)

Auxiliary Power

+5V/0.5A or +8V/0.3A auxiliary output selected by user

Remote ON/OFF Control

External switch or NPN transistor to turn on/off

Power OK Signal

Open drain signal low when PSU turns on. Max. sink current 20mA, max drain voltage 40V

Output Voltage Programming

Adjustment of output voltage between 0-105% of rated output

Output Current Programming

Adjustment of output current between 0-105% of rated output

Operating Temperature Range

-20°C to +60°C. Above 50°C, derate linearly to 60% load at 60°C

Operating Humidity

20-90% RH non-condensing

Storage Temperature Range

-40°C to +85°C

Storage Humidity

10-95% RH non-condensing

Temperature Coefficient

±0.02%/°C (0-50°C)

Vibration

Compliance to IEC60068-2-6-2007, IEC60068-2-64-2008

Safety Standards

UL/cUL (ANSI/AAMI ES 60601-1:2005, 1st edition; CAN/CSA-C22.2 No. 60601-1:08, 2nd edition)
TUV (EN 60601: 2006; IEC 60601-1: 2005); NFPA99: 2005

Withstand Voltage

I/P-O/P: 4KVAC, I/P-FG: 2.0KVAC, O/P-FG: 0.5KVAC

Isolation Resistance

I/P-O/P, I/P-FG, O/P-FG: 100MΩ / 500VDC

EMI Conduction & Radiation

Compliance to EN55011: 2009+A1: 2010

Harmonic Current

Compliance to EN61000-3-2: 2006+A2: 2009 Class A, EN61000-3-3 : 2008

EMS Immunity

Compliance to EN60601-1-2: 2007; IEC61000-4-2: 2008; IEC61000-4-3: 2010; IEC61000-4-4: 2011;
IEC61000-4-5: 2005; IEC61000-4-6: 2008; IEC61000-4-8: 2009; IEC61000-4-11: 2004

Cooling

Internal fan controlled by power rating & temperature, allow 50mm clearance at each end for airflow

Dimensions

267(L) x 127(W) x 63.5(H) mm

Weight

2.6Kg

Electrical Specification

Notes:
1. All parameters NOT specifically mentioned are measured at 230Vac input, rated load and 25°C ambient temperature
2. Ripple and noise are measured at 20MHz bandwidth by using a 12” twisted pair wire terminated with a 0.1µF ceramic & 47µF electrolytic capacitors across the output
3. Tolerance includes set up tolerance, line regulation and load regulation
4. Derating is required at low input voltages. Below 100Vac, derate linearly to 90% load at 90Vac
5. When parallel connected, only one unit may operate if the total output load is less than 5% of rated load
6. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets EMC Directives

PME1200 Series
1200W Active PFC
Single Output

Features

Universal AC Input with active PFC

Programmable Output Voltage (0%-105%)

Programmable Output Current (0%-105%)

Medical Safety Approval EN60601-1, UL60601-1

High Efficiency up to 93%

+5V/0.5A or +8V/0.3A Auxiliary Output

Intelligent LED Indicators

Forced current sharing for parallel operation

Power OK signal (Power good, logic low)

Remote ON/OFF, Remote Sense function

OVP, OLP, OTP, Fan Failure Protections

Built-in I

2

C Serial Data Bus