Powersolve PME1200 Series User Manual
Ac-dc converter, Features
Powersolve Electronics Ltd. 8A Arnhem Road, Newbury, RG14 5RU. England
Tel: 01635 521858 Fax: 01635 523771 Email: [email protected]
www.powersolve.co.uk
AC-DC Converter
Input Voltage
90-264VAC full range, 47-63Hz / 127-370VDC. Below 100VAC, derate load linearly to 90% at 90VAC.
Power Factor
0.95 at 230VAC / 0.99 at 115VAC at full load
AC Input Current (typ.)
14.5A at 115VAC / 6A at 230VAC
Inrush Current
30A at 115VAC / 45A at 230VAC
Leakage Current
<300µA at 264VAC for Earth / <100µA at 264VAC for Patient
Output Voltage
See Table
Output Current
See Table
Voltage Tolerance
±2%
Voltage Adjust Range
±5.0% typical adjustment by potentiometer
Overload Protection
>105% of rated output power. Protection type: Constant current limiting
Overvoltage Protection
Variable OVP, 125% ±7% of Vout. Output latches off, recycle AC input to recover or inhibit
Over Temperature Protection
Shuts down output, recovers automatically (100°C ±5°C detect on heatsink on secondary side)
Auxiliary Power
+5V/0.5A or +8V/0.3A auxiliary output selected by user
Remote ON/OFF Control
External switch or NPN transistor to turn on/off
Power OK Signal
Open drain signal low when PSU turns on. Max. sink current 20mA, max drain voltage 40V
Output Voltage Programming
Adjustment of output voltage between 0-105% of rated output
Output Current Programming
Adjustment of output current between 0-105% of rated output
Operating Temperature Range
-20°C to +60°C. Above 50°C, derate linearly to 60% load at 60°C
Operating Humidity
20-90% RH non-condensing
Storage Temperature Range
-40°C to +85°C
Storage Humidity
10-95% RH non-condensing
Temperature Coefficient
±0.02%/°C (0-50°C)
Vibration
Compliance to IEC60068-2-6-2007, IEC60068-2-64-2008
Safety Standards
UL/cUL (ANSI/AAMI ES 60601-1:2005, 1st edition; CAN/CSA-C22.2 No. 60601-1:08, 2nd edition)
TUV (EN 60601: 2006; IEC 60601-1: 2005); NFPA99: 2005
Withstand Voltage
I/P-O/P: 4KVAC, I/P-FG: 2.0KVAC, O/P-FG: 0.5KVAC
Isolation Resistance
I/P-O/P, I/P-FG, O/P-FG: 100MΩ / 500VDC
EMI Conduction & Radiation
Compliance to EN55011: 2009+A1: 2010
Harmonic Current
Compliance to EN61000-3-2: 2006+A2: 2009 Class A, EN61000-3-3 : 2008
EMS Immunity
Compliance to EN60601-1-2: 2007; IEC61000-4-2: 2008; IEC61000-4-3: 2010; IEC61000-4-4: 2011;
IEC61000-4-5: 2005; IEC61000-4-6: 2008; IEC61000-4-8: 2009; IEC61000-4-11: 2004
Cooling
Internal fan controlled by power rating & temperature, allow 50mm clearance at each end for airflow
Dimensions
267(L) x 127(W) x 63.5(H) mm
Weight
2.6Kg
Electrical Specification
Notes:
1. All parameters NOT specifically mentioned are measured at 230Vac input, rated load and 25°C ambient temperature
2. Ripple and noise are measured at 20MHz bandwidth by using a 12” twisted pair wire terminated with a 0.1µF ceramic & 47µF electrolytic capacitors across the output
3. Tolerance includes set up tolerance, line regulation and load regulation
4. Derating is required at low input voltages. Below 100Vac, derate linearly to 90% load at 90Vac
5. When parallel connected, only one unit may operate if the total output load is less than 5% of rated load
6. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets EMC Directives
PME1200 Series
1200W Active PFC
Single Output
Features
•
Universal AC Input with active PFC
•
Programmable Output Voltage (0%-105%)
•
Programmable Output Current (0%-105%)
•
Medical Safety Approval EN60601-1, UL60601-1
•
High Efficiency up to 93%
•
+5V/0.5A or +8V/0.3A Auxiliary Output
•
Intelligent LED Indicators
•
Forced current sharing for parallel operation
•
Power OK signal (Power good, logic low)
•
Remote ON/OFF, Remote Sense function
•
OVP, OLP, OTP, Fan Failure Protections
•
Built-in I
2
C Serial Data Bus