beautypg.com

Accessories and options - continued – GE P&W Ultrapure and Drinking Water TOC Analyzers - Sievers 5310 C Series TOC Analyzers User Manual

Page 3

background image

GE Autosampler + DataPro 5310 C* Software
(Laboratory and Portable only)

For high-volume laboratory appli-
cations, the 5310 C can be used
with the GE Autosampler. The GE
Autosampler provides high-
throughput sample capacity (up to
63 positions for 40- or 60-mL vials,
or up to 120 positions for 35-mL
test tubes or 17-mL vials). The
Sievers DataPro 5310 C software
integrates the GE Autosampler
with the 5310 C Laboratory and Portable Analyzers to
offer a host of productivity-enhancing features, such as
automated calibration and verification protocols.
DataPro 5310 C also gives users full sampling flexibility
with custom sample protocols and user-defined sam-

pling capabilities. The GE Autosampler’s slim 28.2-cm
(11.1-in) profile requires minimal bench space.

In addition to the GE Autosampler’s vial positions, a
standards rack provides an additional six vial positions
that can be used to check standards that are typically
run with each batch of samples. Through the 5310 C
DataPro software, the Autosampler can be programmed
to perform a cal check between samples, at a frequency
determined by the user. Alternatively, the standards
rack can be used for additional sample vial capacity.

The GE Autosampler also has an optional rinse station
and wash pump that provides a flush of the sample
inlet system with a flowing water source. This
maximizes analytical performance of sample sets with
widely varying concentration.

Accessories and Options - continued

Recommended System Requirements for DataPro 5310 C Software

System Component

Recommended Requirements

Operating System

Windows 2000/XP

Windows/Vista and Windows 7

RAM

256 MB

1 GB

CPU

500 MHz

1 GHz 32-bit (x86) or 64-bit (x64)

Screen Resolution

1024 x 768

1024 x 768

Screen Color

65K (high color)

32-bit (True Color)

Mouse/Pointing Device

Optional

Serial Port

Required to connect to the Analyzer

Printer

Optional to run program; required to print protocols and results

CD-ROM Drive

Required for installation

GE Autosampler Specifications

Capacity

Configuration options include a six-position 40-mL standards rack and: Sixty three (63)
40-mL or 60-mL vials, or one hundred and twenty (120) 17-mL vials or 35-mL tubes

Power

60W (base model), 100-240 VAC, 50-60 Hz. No manual voltage adjustments are required.
The power supply unit has the following approvals: CE and recognized cRU component
for Canada and US.

Dimensions

H: 52.2 cm (20.5 in); W: 29.2 cm (11.5 in); D: 53.3 cm (21.0 in)

Weight

15 kg (33 lbs)

Safety Certifications

CE, ETL