VEGA VEGAMET 514 User Manual
Page 8
8
VEGAMET 514
Ambient conditions
Permissible ambient temperature
-20°C … +60°C
Storage and transport temperature
-40°C … +80°C
Electrical connection
Multiple plug component
acc. to DIN 41 612, series F, 33-pole, 3 lines
d, b, z (partly equipped)
Carrier BGT 596 (Ex)
connection to the appropriate module
Housing type 505
connection to screw terminals (max. 2.5 mm
2
)
Electrical protective measures
Protection
- unassembled
IP 00
- mounted into
carrier BGT 596
- front side completely equipped
IP 40
- upper and lower side
IP 20
- wiring side
IP 00
- mounted into
housing type 505 or type 506
- terminal side
IP 20
- housing general
IP 30
Protection class
II (in housing type 505)
Overvoltage category
II
Electrical separating measures
Reliable separation acc. to
VDE 0106, part 1 between
power supply, fail safe and level relay
and measuring data inputs
- reference voltage
250 V
- isolation resistance
2.3 kV
Galvanic isolation
between the relay outputs
- reference voltage
250 V
- isolation resistance
1.4 kV
Potential separation
between DISBUS and outputs
- reference voltage
50 V
- isolation resistance
0.5 kV
Common reference potential at
voltage and current output,
correction signal output
Mechanical data
Series
module unit for carrier BGT 596
or housing type 505
Dimensions
B = 25.4 mm (5 TE), H = 128.4 mm, T = 162 mm
CE conformity
VEGAMET 514 and VEGAMET 514 Ex signal conditioning instruments meet the protective
regulations of EMC (89/336/EWG) and NSR (73/23/EWG). Conformity has been judged
acc. to the following standards:
EMC
Susceptibility
EN 50 081 - 2: 1993
Immission
EN 50 082 - 2: 1995
NSR
EN 61 010 - 1: 1993
Product description