9 simulation, 11 password, language, menu mode, 10 special function: reset – VEGA VEGAMET 514 User Manual
Page 44: 9 simulation 7.10 special function: reset, See chapter "6 settings in the reduced menu, Settings in the "extended menu
44
VEGAMET 514
7.11
Password, language, menu mode
See chapter "6 Settings in the reduced menu".
Param.
TAG-
No. 1
Confi-
gura-
tion
Add'l
func-
tions
Pass-
word
off
TAG -
No. 1
%
Info
Edit
lin.
curve
Lan-
guage
Eng-
lish
Reset
VEGA-
MET
Ser-
vice
Menu
mode
Ex-
tended
Redu-
ced
Settings in the "extended menu"
7.9 Simulation
7.10
Special function: Reset
To check the outputs and connected instru-
ments, you can adjust in this menu item any
individual percentage value with the keys [+]
and [–]. Initial point is always the actual
measured value. The indicated value flashes
when the simulation is activated. 15 minutes
after setting the simulated value, the
simulation is automatically terminated and the
measured value indication switched back on.
Param.
TAG-
No. 1
Adjust-
ment
Signal
condi-
tioning
Out-
puts
Simu-
lation
Simu-
lation
%
XX,X
Simu-
lation
Now!
OK?
TAG -
No. 1
%
xx,x
Special
func-
tions
Reset
Level
Reset
OK ?
Reset
Now!
OK ?
Param.
TAG-
No. 1
Adjust-
ment
Signal
condi-
tioning
Out-
puts
Simu-
lation
TAG -
No. 1
%
xx,x
With this reset, all parameter values of the
measurement loop TAG-No. 1 are reset to
factory setting (empty and full adjustment,
relay outputs, current output, volt output and
scaling of the meas. value indication are
reset).