D. measurement techniques – Veris Industries 133 Manual User Manual

Page 6

background image

10

2.

Step by Step Procedures:

a.

Measuring DC Volts

CAUTION!
Do not attempt to make a voltage measurement if
a test lead is plugged in the A or µmA input jack.
Instrument damage and/or personal injury may result.

WARNING!

Do not attempt to make a voltage measurement of more
than 1000V or of a voltage level that is unknown.

Instrument set-up:

FUNCTION

BLACK

RED

MINIMUM

MAXIMUM

TEST LEAD

TEST LEAD

READING

READING

DCV

COM

V

0.1mV

1000V

Measurement Procedure:

1.

Disconnect power to circuit to be measured.

2.

Plug black test lead into the COM input jack.

3

Plug red test lead into V input jack.

4.

Set the rotary switch on the 133 to the desired range
in the DCV function depending on the voltage to be
measured.

5.

Connect test leads to circuit to be measured.

6.

Reconnect power to circuit to be measured.

7.

Read the voltage on the 133.

Application Notes

When measuring DC Voltage of a battery, the most accu-
rate reading can be attained by testing the battery under
load. To accomplish this, follow steps 1 through 4 above
and the following (with the battery in holder and device
turned on):

Connect the red test lead from the meter to the

positive (+) terminal of the battery.

Connect the black test lead to the negative (-) terminal

of the battery.

Reconnect power to the circuit and read the voltage on

the 133.

D. MEASUREMENT TECHNIQUES

1.

Controls and Functions:

Push Buttons

Turns the 133 on and off.

Data-H

Activates the Data Hold function.

Rotary Switch
DCV

Used for measurement of DC Volts. Select the
best range for the voltage to be measured.

ACV

Used for measurement of AC Volts. Select the
best range for the voltage to be measured.

DCA

Used for measurement of DC Amps. Select the
best range for the current to be measured.

ACA

Used for measurement of AC Amps. Select the
best range for the current to be measured.

OHM

Used for measurement of Resistance, Diode Test
and Continuity Buzzer. Select the best range for
the resistance to be measured.

Input Jacks
A

Red test lead connection for current measure-
ments on the 2 and 10 ACA and DCA functions.

mAµA

Red test lead connection for current measure-
ment on the mA and µA DCA and ACA functions.

COM

Black test lead connection for all functions.

V

Red test lead connection for OHM, DCV and ACV
functions.

11

!