Veris Industries TK SERIES Install User Manual
Tk series, Temperature, Installation guide
Installation Guide
Temperature
Z205789-0C
Page 1 of 2
©2013 Veris Industries USA 800.354.8556 or +1.503.598.4564 / [email protected] 08131
Alta Labs, Enercept, Enspector, Hawkeye, Trustat, Aerospond, Veris, and the Veris ‘V’ logo are trademarks or registered trademarks of Veris Industries, L.L.C. in the USA and/or other countries.
Other companies’ trademarks are hereby acknowledged to belong to their respective owners.
TM
NOTICE
• This product is not intended for life or safety applications.
• Do not install this product in hazardous or classified locations.
• Read and understand the instructions before installing
this product.
• Turn off all power supplying equipment before working on it.
• The installer is responsible for conformance to all applicable codes.
No responsibility is assumed by Veris Industries for any consequences arising out of the
use of this material.
Product Identification
Product Overview
TK ceiling mounted temperature sensors monitor the temperature of the air in a duct. The design includes a
pre-calibrated sensor housed in a sturdy stainless steel probe with no junction box. The TK has a resistive output
compatible with building control systems. It is warranted to meet accuracy specifications for a period of five years.
TK Series
Temperature Probe
Specifications
Wiring 22 AWG; 2-wire RTD/Thermistor
Housing Stainless Steel
LINITEMP OPTION
Input Power 5-30 VDC
Output 1 µA/°C or 10 mV/°C
Operating Temperature
Range
-25° to 105°C (-13 to 221°F)
Accuracy
Calibration Error
Error over Temperature
1.5°C (35°F) typical; 2.5°C (37°F) max. at 25°C (77°F)
1.8°C (35°F) typical; 3.0°C (34°F) max. over 0° to 70°C (32° to 158°F) range;
2.0°C (35°F) typical; 3.5°C (38°F) max. over -25° to 105°C (-13° to 221°F) range
Increased cable length affects the readings of lower resistance RTDs (100R platinum RTD).
Room temperature error is documented on each unit.
Sensor Type
Probe Length
B = 100R platinum, RTD
C = 1k platinum, RTD
D = 10k T2, Thermistor
E = 2.2k, Thermistor
F = 3k, Thermistor
G = 10k CPC, Thermistor
H = 10k T3, Thermistor
I = 1k Balco (Nickel-iron) RTD
J = 10k Dale, Thermistor
K = 10k w/11k shunt, Thermistor
M = 20k NTC, Thermistor
N = 1800 ohm, Thermistor
P = 10mV/°C, Linitemp
R = 10k US, Thermistor
S = 10k 3A221, Thermistor
T = 100k, Thermistor
U = 20k “D”, Thermistor
CC = 15k, Thermistor
B = 4 1/2" (115 mm)
C = 6” (153 mm)
D = 8" (204 mm)
R = 14” (356 mm)
TK
= Resistive
Cal Certs
0 = None
1 = 1 Point Cal Cert
2 = 2 Point Cal Cert
R0
Output