Veris Industries 621 Manual User Manual

Page 5

background image

Step by Step Procedures

1. Press the 'ON' key and the LCD display will be

activated and automatically zeroed (if it does not
read 0.00 proceed as follows

2. Momentarily Press the key and the display

should zero within +/- 0.05 if not repeat until it
does

3. Connect pressure meter to installation or

equipment to be tested using appropriate hose
connection in accordance with the manufacturers
instructions

4. The pressure reading will appear on the display.

To activate the backlight in poor light conditions
press the key at any time

5. If the reading is erratic or changing activate the

smoothing function by pressing the '=' key

6. To convert the displayed pressure reading to

another engineering unit depress the 'UNIT' key
anytime and scroll through to the desired units of
pressure.

Accessories & Specifications

Standard accessories

9V battery . . . . . . . . . . . . . . . . . . . . . .A009
Soft Pouch . . . . . . . . . . . . . . . . . . . . . .A600

Optional Accessories

2 metre (6 ft) nitrile rubber tubing . . . .A601
Large Zipped pouch . . . . . . . . . . . . . . .A904

Specifications

Auto Range

0.0 to -300.0 mbar
0.0 to 120.4 inH2O

Accuracy

±(0.2% of F.V.)

Over Pressure

300 mbar

Power Supply

9V

Relative Humidity

0% to 80%

Operating Temp.

0ºC to +40ºC

Storage Temp.

-10ºC to +50ºC

Auto Power Off

40 Minutes

Input Type

Differential