beautypg.com

Veris Technologies pH Manager - Operating Instructions User Manual

Page 38

background image

Pub.# OM17-MSP-pH

4-9


2. Adjust electrode holder: with sampling mechanism raised completely, adjust electrode

holder to provide ½” (1.2 cm) clearance between it and sampling trough.

Figures 15 and 16

3. Wash adjustment: Wash brackets should be parallel to sampling trough, with jets directly

beside electrode holde

r, jets should be 1” (2.5 cm) below electrodes; when electrodes are

properly aligned.

Figures 17a and 17b

Figures 18a and 18b

Turn crank to raise
or lower electrode
holder

electrode
holder 1”
(2.5cm)
above wash
jets

When wash jets are
properly aligned,
overspray is minimized
and water bubbles out
top of empty electrode
holder as shown here.

½” (1.2 cm) clearance
between electrode holder
and trough liner

When installing BNC cover, route
electrode wires under box; center box
on white pad, and tighten wingnut finger
tight. Keep cover installed even when
electrodes are removed.