Veris Technologies pH Manager - Operating Instructions User Manual
Page 38
Pub.# OM17-MSP-pH
4-9
2. Adjust electrode holder: with sampling mechanism raised completely, adjust electrode
holder to provide ½” (1.2 cm) clearance between it and sampling trough.
Figures 15 and 16
3. Wash adjustment: Wash brackets should be parallel to sampling trough, with jets directly
beside electrode holde
r, jets should be 1” (2.5 cm) below electrodes; when electrodes are
properly aligned.
Figures 17a and 17b
Figures 18a and 18b
Turn crank to raise
or lower electrode
holder
electrode
holder 1”
(2.5cm)
above wash
jets
When wash jets are
properly aligned,
overspray is minimized
and water bubbles out
top of empty electrode
holder as shown here.
½” (1.2 cm) clearance
between electrode holder
and trough liner
When installing BNC cover, route
electrode wires under box; center box
on white pad, and tighten wingnut finger
tight. Keep cover installed even when
electrodes are removed.