beautypg.com

Veris Technologies pH Manager - Operating Instructions User Manual

Page 11

background image

Pub.# OM17-MSP-pH

3-3

Repeat for

electrode 2’s ID and press 3 to confirm.


The instrument will prompt for the electrodes to be inserted into pH buffer 4 solution; Slide cup with
pH 4 buffer solution onto electrode holder. Press 1 to continue with calibration or 2 to exit. Tips:
Don’t overfill solution. Cup only needs enough solution to immerse electrode tip and face. Don’t
reuse solutions.

Figure 9

Figure 10

The instrument will read the electrodes for 10 seconds, displaying the output (as it counts
seconds):

Figure 11

After 10 seconds, the instrument will display the final pH reading and offer the options to 1) Accept
pH 4 buffer readings; 2) Redo pH 4 buffer readings; or 3) Exit pH electrode calibration
. If the
readings are satisfactory, press 1; if the readings are suspect, press 2 to return to re-do pH 4.

Figure 12

After accepting the pH 4 buffer readings, the Instrument will prompt for the electrodes to be
inserted into pH 7 buffer solution. Remove the pH 4 buffer solution cup from the electrode
holder.Rinse the electrodes, electrode holder, and solution cup using the manual wash for at least
10 seconds. Slide the pH buffer 7 solution cup onto the electrode holder.