EN-37

Adventurer Pro

Symptom

Possible Cause

Remedy

Balance will not turn on

• AC power not connected

• Connect AC adapter

• Batteries discharged

• Replace batteries

Battery indicator is flashing

• Batteries are weak

• Replace batteries

Balance does not display

• Improper calibration

• Perform calibration

accurately

• Unstable environment

• Move balance to a

suitable location

Cannot calibrate the balance

• Calibration menu locked

• Turn Calibration menu

lock off

• LFT set to on

• Turn LFT off

• Unstable weight reading

• Eliminate vibration and

drafts.

Cannot change Menu settings

• Submenu locked

• Unlock submenu

• LFT set to on

• Turn LFT off

Error 7.0

• Unstable weight reading

• Eliminate vibration and

when defining the

drafts.

reference weight.

Error 8.1

• Weight reading exceeds

• Clear platform.

Power On Zero limit.

Error 8.2

• Weight reading below

• Put platform on

Power On Zero limit.

balance.

Error 8.3

• Weight reading exceeds

• Clear platform.

overload limit.

Error 8.4

• Weight reading below

• Put platform on

underload limit.

balance.

Error 9.0

• Internal fault.

• Return balance for service.

4. MAINTENANCE
4.1 Calibration

Periodically verify calibration by placing an accurate weight on the balance and if calibration is
required, refer to section 3.7.1.

4.2 Cleaning

If cleaning is required, use a soft cloth dampened with water and a mild detergent. Do not allow
liquids to enter the balance. Do not use harsh chemicals to clean the balance, as the finish may
be damaged.

4.3 Troubleshooting

The following table lists common problems, possible causes and remedies. If the problem persists,
contact Ohaus or your authorized Ohaus dealer.