3.7.1 Calibration

Adventurer Pro balances offer a choice of five calibration methods: Internal Calibration (InCAL

TM

),

Span Calibration, Linearity Calibration, Calibration Test

TM

and Calibration Adjust.

InCal

InCal

InCal

InCal

InCal

TM

- For models equipped with Internal calibration (InCAL

TM

), calibration of the

balance is accomplished by an internal mass.

Span

Span

Span

Span

Span - Span calibration ensures that the balance reads correctly within

specifications using two weight values: zero and a weight value between
25% and 100% of the balance’s capacity.

Linearity

Linearity

Linearity

Linearity

Linearity - Linearity calibration minimizes deviation between actual and displayed

weights within the balance’s weighing range. Three weight values are
used: zero, a weight value at midpoint of the balance's weighing range,
and a weight value at or near the balance’s specified capacity.

Cal Test

Cal Test

Cal Test

Cal Test

Cal Test - Calibration test allows the stored calibration data to be tested against the

current mass being used for the test.

Cal Adj

Cal Adj

Cal Adj

Cal Adj

Cal Adj - For models equipped with InCal

TM

, allows adjustment of the internal

calibration.

Notes: Calibration may be locked out to prevent unauthorized personnel from changing calibration.

if calibration has been locked out, you can only access Cal Test on InCal

TM

models.

Before beginning calibration, have masses available. Masses required for calibration are
listed in the table in section 3.7. 2. Default Span Calibration Points are shown in Bold
print.