Adventurer Pro

EN-26

NOTICE:

Linearity Calibration is required only if the linearity error exceeds tolerance in
specification table.

With the balance on, press and hold the

Menu-Cal button

until MENU CALIBRATE is displayed, then release the button.
CALIBRATE (blinking).

Press the

Yes button to enter calibration.

Note: If the balance has InCal, press the No button to advance
to linearity calibration. CAL LINEARITY (blinking) is displayed.

Press the Yes button and follow the screen instructions. Use
the weight values shown on the primary display.

First the zero reading is taken. Next the display shows the first
calibration weight value. PUT WEIGHT (blinking) is displayed.
Place the specified calibration weight on the platform.

After a few seconds, place the second specified calibration
weight on the plattform. PUT WEIGHT (blinking) is displayed.
After a few seconds, the calibration weight is displayed.

LINEAR DONE is momentarily displayed when calibration is
completed.

After calibration, remove weight from platform. To leave the
menu, press Exit.

To abort calibration at any time, press the Exit button.

Linearity Calibration