Adventurer Pro

EN-28

Calibration Adjust (InCal models)
Calibration Adjust may be used to adjust the result of the internal calibration by ±100 divisions.

Note::::: Before making a calibration adjustment, perform an internal calibration. To verify whether an
adjustment is needed, place a test mass on the platform and note the difference (in divisions)
between the nominal mass value and the actual reading. Refer to Section 3.7.2 and use the
highest value from the Span Calibration Points column as the test mass. If the difference is within
±1 division, calibration adjustment is not required. If the difference exceeds ±1 division, calibration
adjustment is recommended. Following a calibration adjustment, repeat the internal calibration
and verification procedures.

To perform a calibration adjustment, press the

Menu-Cal button

and release it when MENU CALIBRATE is displayed.

Press Yes to enter the Calibration sub-menu. Press No until
CAL ADJ is displayed.

Press Yes to enter the CAL ADJ menu item and view the current
setting. If the actual reading was less than the nominal mass
value, a positive adjustment is required. Press No

until the

setting matches the difference noted earlier in the procedure. If
the actual reading was greater than the nominal mass value, a
negative adjustment is required. Press Back

Back

Back

Back

Back until the setting

matches the difference noted earlier in the procedure. Press Yes
to accept and store the setting.