2 installation, 1 inspecting and unpacking, 2 power supply system mounting – American Magnetics 4Q1005PS-430 Integrated Power Supply System User Manual
Page 29: Installation
Rev.
5
11
2 Installation
Warning
Before energizing the equipment, the earth ground of the power
receptacle must be verified to be at earth potential and able to carry
the rated current of the power circuit. Using extension cords should
be avoided; however, if one must be used, ensure the ground
conductor is intact and capable of carrying the rated current.
In the event that the ground path becomes less than sufficient to
carry the rated current of the power circuit, the equipment should be
disconnected from power, labeled as unsafe, and removed from place
of operation.
Do not operate this equipment in the presence of flammable gases.
Doing so could result in a life-threatening explosion.
Do not modify this equipment in any way. If component replacement
is required, return the equipment to AMI facilities as described in
the Troubleshooting section of this manual.
If used in a manner not specified in this manual, the protection
provided by the design, manufacture and documentation of the
system may be impaired.
2.1 Inspecting and Unpacking
Carefully remove the equipment, interconnecting cabling, and
documentation CD (and/or printed material) from the shipping carton, and
remove all packaging material.
Note
If there is any shipping damage, save all packing material and
contact the shipping representative to file a damage claim. Do not
return to AMI unless prior authorization has been received.
2.2 Power Supply System Mounting
If the system is to be used on a table top, place the equipment on a flat,
secure surface.