3 setup configuration commands and queries, Remote interface reference – American Magnetics 4Q1005PS-430 Integrated Power Supply System User Manual
Page 128
110
Rev. 5
Remote Interface Reference
SETUP Configuration Commands and Queries
•
*CLS
Clears the Standard Event register and the error buffer.
•
*ESR?
Returns a decimal sum which corresponds to the binary-weighted sum of
the contents of the Standard Event register.
•
*ESE <
enable_value
>
Enables bits in the Standard Event register to be reported in the
“Standard Event” bit (bit 5) of the Status Byte register. To enable bits, you
must write a decimal
<
enable_value
>
which corresponds to the binary-
weighted sum of the bits you wish to enable. Refer to Table 4-2 on page 104
for more information. For example, to enable all categories of error
messages to be reported in bit 5 of the Status Byte register, send:
*ESE 60;
•
*ESE?
The
*ESE?
query returns a decimal sum which corresponds to the binary-
weighted sum of the bits enabled by the last
*ESE
command.
•
*PSC {0|1}
Power-On Status Clear. If
*PSC 1
is in effect, the Standard Event enable
register and the Status Byte enable register are cleared at power on. If
*PSC 0
is in effect, the enable registers are not cleared at power on. The
default setting is “1”.
•
*PSC?
Returns the Power-On Status Clear setting currently in effect. A value of
“0” indicates the enable registers are not cleared at power on; a value of “1”
indicates the enable registers are cleared at power on.
•
*OPC
Sets the “Operation Complete” bit (bit 0) of the Standard Event register
when executed. See page 104 for a complete discussion.
•
*OPC?
Returns “1” to the requesting interface when executed. See page 105 for
more information.
4.5.3
SETUP Configuration Commands and Queries
The SETUP Configuration Commands and Queries provide read/write
access to the setup functions available within the Supply (see section
3.10.1), Load (see section 3.10.2), and Misc (see section 3.10.3) submenus.