Remote interface reference – American Magnetics 4Q1005PS-430 Integrated Power Supply System User Manual
Page 137
Rev. 5
119
Remote Interface Reference
SETUP Configuration Commands and Queries
•
CONFigure:LOCK:COILconst {0|1}
Specifies whether the Coil Constant value (under the Load submenu) is
locked or unlocked. Sending “0” unlocks. A “1” locks. “0” is the default
value.
•
LOCK:VOLTage:COILconst?
Returns “0” for Coil Constant value (under the Load submenu) unlocked,
or “1” for locked.
•
CONFigure:LOCK:CURRent:LIMit {0|1}
Specifies whether the Current Limit value (under the Load submenu) is
locked or unlocked. Sending “0” unlocks. A “1” locks. “0” is the default
value.
•
LOCK:CURRent:LIMit?
Returns “0” for Current Limit value (under the Load submenu) unlocked,
or “1” for locked.
•
CONFigure:LOCK:PSwitch:SETtings {0|1}
Specifies whether persistent switch settings are locked or unlocked.
Persistent switch settings protected by this setting (all under the Load
submenu) are: the PSwitch Installed picklist value, the PSwitch Current
value, the PSwitch Heated Time value, PSwitch Current Detect, PSwitch
Cooled time, PSw P/S Ramp Rate, and PSwitch Cooling Gain. Sending “0”
unlocks. A “1” locks. “0” is the default value.
•
LOCK:PSwitch:SETtings?
Returns “0” for persistent switch settings unlocked, or “1” for locked.
•
CONFigure:LOCK:QUench:DETect {0|1}
Specifies whether the Enable Quench Detect picklist value (under the
Load submenu) is locked or unlocked. Sending “0” unlocks. A “1” locks. “0”
is the default value.
•
LOCK:QUench:DETect?
Returns “0” for Enable Quench Detect picklist value (under the Load
submenu) unlocked, or “1” for locked.
•
CONFigure:LOCK:ABsorber {0|1}
Specifies whether the Energy Absorber Present picklist value (under the
Load submenu) is locked or unlocked. Sending “0” unlocks. A “1” locks. “0”
is the default value.