beautypg.com

Set v, i, and p limits triggered setpoints – Xantrex Technology GPIB-M-XT User Manual

Page 67

background image

Operation

Auto Sequencing

Release 2.1

65

Set V, I, and P

Limits

The voltage, current and power setpoints can be limited to less than the supply rating
range to match the tolerance of connected equipment or any other criteria you may
have.

Once the limits have been changed from the supply’s default rated output, settings
outside this range are no longer accepted.

To set the voltage limits:
SOUR:VOLT:LIM:HIGH
SOUR:VOLT:LIM:LOW

To set the current limits:
SOUR:CURR:LIM:HIGH
SOUR:CURR:LIM:LOW

If the high end of the range was set to 5 volts, the command, SOUR:VOLT 10 would
return an error.

Triggered

Setpoints

To set a triggered setpoint:
SOUR:VOLT:TRIG
SOUR:CURR:TRIG

See

“Triggering Commands” on page 66

for more information.

Note Setpoint limits do not apply to the triggered outputs and auto sequence outputs.

This manual is related to the following products: