beautypg.com

Exit calibration mode restore factory calibration – Xantrex Technology GPIB-M-XT User Manual

Page 131

background image

Calibration

Exit calibration mode

Release 2.1

129

Exit calibration mode

When you have completed calibration, exit calibration mode by sending the
command:

CAL:STAT OFF, "0000"

where “0000” is replaced with your security code.

Restore Factory Calibration

Restore the unit to the calibration constants set at the factory by sending the
command:

CALibration:RESTore

Note This procedure should not be used in place of regular calibration, but may be useful to

restore the unit to an operational state in case of failure.

This manual is related to the following products: