Configure output protection – Xantrex Technology GPIB-M-XT User Manual
Page 53
Operation
Power Supply Operation
Release 2.1
51
Configure
Output
Protection
Five configurable protection mechanisms are available:
•
OVP: Over-Voltage Protection. Factory default = 0V (disabled)
•
UVP: Under-Voltage Protection. Factory default = 0V (disabled)
•
OCP: Over-Current Protection. Factory default = 0A (disabled)
•
UCP: Under-Current Protection. Factory default = 0W (disabled)
•
Fold: Fold Protection. (See below.)
OVP shuts down the power supply if the protection limit is exceeded. The other
options offer a choice: they shut down the power supply or issue a warning by setting
a bit in the status register. Set the protection mechanism to ON to enable it to
shutdown the power supply.
When the protection level is set to zero, that mechanism is disabled. However, in the
case of OVP, a fixed hardware protection mechanism still exists to protect the supply
from damage.
To set the over-voltage protection level:
SOUR:VOLT:PROT
To set the under-voltage protection level:
SOUR:VOLT:PROT:UND
To enable UVP shutdown:
SOUR:VOLT:PROT:UND:STAT
To set the over-current protection level:
SOUR:CURR:PROT
To enable OCP shutdown:
SOUR:CURR:PROT:STAT
To set the under-current protection level:
SOUR:CURR:PROT:UND
To enable UCP shutdown:
SOUR:CURR:PROT:UND:STAT
To check if a protection mechanism was tripped:
SOUR:VOLT:PROT:OVER:TRIP?
SOUR:VOLT:PROT:UND:TRIP?
SOUR:CURR:PROT:OVER:TRIP?
SOUR:CURR:PROT:UND:TRIP?
Alternatively, you can query the status register. See
.