beautypg.com

Calibration procedure, Output voltage – Xantrex Technology GPIB-M-XT User Manual

Page 129

background image

Calibration

Calibration Procedure

Release 2.1

127

Calibration Procedure

Calibration of programming and readback are combined in a single procedure.

Output

Voltage

1. Set the load to open circuit. Attach a DVM across the output terminals.
2. Minimum calibration level

Set the output voltage to 10% by sending the command:

CAL:OUTP:VOLT:LEV MIN

3. Enter voltage data

Enter the voltage read from the external DVM.

CAL:OUTP:VOLT:DATA

4. Maximum calibration level

Set the output voltage to 90% by sending the command:

CAL:OUTP:VOLT:LEV MAX

5. Enter voltage data

Enter the voltage read from the external DVM.

CAL:OUTP:VOLT:DATA

6. Power supply calculates and stores calibration constants.

This manual is related to the following products: