beautypg.com

Table 1.4 – Xantrex Technology GPIB-M-XT User Manual

Page 24

background image

Features and Specifications
Specifications

22

Operating Manual for Multichannel Functionality (GPIB-M)

Table 1.4 Specifications for XHR 1000 W with GPIB-M or CANbus

Models

7.5-130

20-50

33-33

40-25

60-18

Program Resolution

Voltage
Current

1.16mV
42.0mA

1.8mV
30.8mA

3.08mV
18.2mA

6.2mV
9.8mA

9.2mV
6.44mA

Program Accuracy

1

Voltage

Current

10mV
±0.12%
900mA
±0.1%

50mV
±0.12%
750mA
±0.1%

75mV
±0.12%
500mA
±0.1%

75mV
±0.3%
350mA
±0.1%

150mV
±0.25%
250mA
±0.1%

Readback Resolution

Voltage
Current

1.16mV
42.0mA

1.8mV
30.8mA

3.08mV
18.2mA

6.2mV
9.8mA

9.2mV
6.44mA

Readback Accuracy

1

Voltage

Current

30mV
±0.12%
900mA
±0.1%

60mV
±0.12%
750mA
±0.1%

75mV
±0.12%
500mA
±0.1%

75mV
±0.3%
350mA
±0.1%

150mV
±0.25%
250mA
±0.1%

Models

100-10

150-7

300-3.5

600-1.7

Program Resolution

Voltage
Current

15.4mV
3.92mA

23.1mV
2.52mA

46.2mV
1.26mA

92.4mV
0.56mA

Program Accuracy

1

Voltage

Current

1.

Apply accuracy specifications according to the following voltage program accuracy example:

Set a model XHR 20-50 power supply to 10 volts.
The expected result will be within the range of 10 volts ± 50mV ± 0.12% of the set voltage of 10 volts.

150mV
±0.35%
140mA
±0.15%

225mV
±0.35%
120mA
±0.1%

225mV
±0.35%
80mA
±0.1%

250mV
±0.35%
80mA
±0.1%

Readback Resolution

Voltage
Current

15.4mV
3.92mA

23.1mV
2.52mA

46.2mV
1.26mA

92.4mV
0.56mA

Readback Accuracy

1

Voltage

Current

150mV
±0.35%
140mA
±0.15%

225mV
±0.35%
120mA
±0.1%

225mV
±0.35%
80mA
±0.1%

250mV
±0.35%
80mA
±0.1%

This manual is related to the following products: