Chapter 2

Connecting Signals

SCXI-1503 User Manual

2-4

ni.com

2-Wire Configuration

The basic

2-wire configuration is shown in Figure 2-2. In this configuration

an error voltage (V

E

) is introduced into the measurement equal to the

excitation current (I

EX

) times the sum of the two lead resistances, R

L1

and

R

L2

. If we assume equal lead resistances, R

L1

= R

L2

= R

L

, the magnitude of

the error voltage is:

This is the most commonly used configuration for connecting thermistors
to a signal conditioning system because the large sensitivity of thermistors
usually results in the introduction of a negligible error by the lead
resistances.

RTDs typically have a much smaller sensitivity and nominal resistance than
thermistors, therefore a 2-wire configuration usually results in the
introduction of larger errors by the lead resistance.

Figure 2-2. 2-Wire Resistive Sensor Connected in a 2-Wire Configuration

V

E

2R

L

I

EX

=

ON

CH X

+

SCXI-1306

External Sensor

SCXI-1503

Channel X

IEX+

AI+

AI–

IEX–

I = 100 µA

R

L1

R

L2

R

T