Precision Trading 60 R3.0 User Manual
Page 46
![background image](/manuals/176805/46/background.png)
Operator Manual, Supercharger 60
R3.0
Page 46 of 62
• Set the Cell Selector at 20, connect a voltage source set to
approximately 21V and start the Charger-Analyzer in AUTO
Discharge
•
Adjust for a Discharge CUT-OFF (Capacity Failure) at 20.0 volts.
• NOTE: This step requires that the Overvoltage cut-off on the
MONITOR circuit board be previously calibrated.
9.2.3.3. CLOCK-TIMER
• No adjustments on this board.
• Older versions of the Charger-Analyzer use two circuit boards
associated with timer functions: The Timer Board 9879903010 and
the Clock Display Board 9879903009.
• Newer Superserseders use a single Clock-Timer Board
9879903109. In either case, there are no adjustments to perform.
9.2.3.4. (Blank)
9.2.3.5. CURRENT CONTROL (see fig 11).
NOTE: Use the hook at pin 14 as ground reference for voltage
measurements (test point E0).
9.2.3.5.1.
R29 - 10 VOLT ADJUSTMENT.
• Static test (no battery or Calibrator).
• Adjust for a reading of 10.000V ±0.003V at E7.
• NOTE: Newer versions of this circuit board have a
fixed 10V reference (no adjustment). Verify that it
reads 10V ±0.005V.
9.2.3.5.2.
R28 - LOW CURRENT CALIBRATE.
• Connect a battery and start the Charger-Analyzer in
Single Rate.
• Adjust for a reading of 0.9 to 1.0 AMP at the current
digital panel meter, with the topping charge current
potentiometer set for 1 AMP (010).
9.2.3.5.3.
R25 - HIGH CURRENT CALIBRATE
• Connect a battery and start the Charger-Analyzer in
Dual Rate.
• Adjust for a reading of 25.0 AMPS at the current
digital panel meter, with the main current
potentiometer set for 25.0 AMPS (250).