Precision Trading 60 R3.0 User Manual
Page 30
Operator Manual, Supercharger 60
R3.0
Page 30 of 62
6.5 DISCHARGE
NOTE: Do not discharge batteries on the Temp-Plate. Normal battery heating of the
Temp-Plate during discharge will generate a false battery Overtemp fault when
attempting to recharge it.
6.5.1. Current and Time
6.5.1.1. Program the discharge rate and time duration.
6.5.1.2. Verify that the Timer Speed is consistent with the required elapsed
time.
6.5.2. RUN
6.5.2.1. Depress auto discharge (BLUE) control button, for analysis to one volt
per cell.
6.5.2.2. Depress full discharge (RED) control button for battery deep cycling.
6.5.2.3. DISCHARGE will terminate automatically (CYCLE END) when the
(TOTAL) time selected is reached or (CAPACITY FAILURE) if the
BATTERY voltage reaches the equivalent of ONE VOLT per CELL
prior to the selected time.
6.5.2.4. NOTE: Battery voltage is ignored during full discharge.
6.5.2.5. Discharge will terminate as a FAULT if the MONITOR CIRCUIT
detects:
• An overheating of the discharge transistors (OVERHEAT
FAULT).
• That the discharge current differs from the programmed value by
more than one AMP (CURRENT FAULT).
• That the programmed current exceeds 30A for a battery voltage in
excess of 32V.