2.3 Peak Voltage Charge
2.3.1. Nominal Battery Voltage
2.3.1.1. Enter the nominal battery voltage using the cell selector (see chart on
page
xxx
).
2.3.1.2. Select transfer from Main to Topping on Peak or stop on Peak.
2.3.1.3. Program the Main and Topping Charge currents as required.
2.3.1.4. Program the Main and Total Time selectors to allow the battery to
reach the required charge level (consult battery manufacturers
specifications). Verify that the Timer Speed is consistent with the
required elapsed time.
2.3.1.5. Start the Charger-Analyzer in the Dual Rate mode. The charger will
transfer from Main to Topping charge when the battery reaches the
peak voltage or Start in the Dual or Single Rate mode and the Charger-
Analyzer will stop on peak (at any part of the cycle).
2.3.1.6. The charge will terminate automatically (CYCLE END) when the
TOTAL TIME selected is reached.
2.3.1.7. The charge will terminate as a FAULT if the MONITOR CIRCUIT
detects that:
• Battery temperature exceeds 45
o
C/113
o
F (BATTERY
OVERTEMP FAULT).
• The actual charge current deviates from the programmed value by
more than one AMP (CURRENT FAULT).