Fluke 5020A User Manual
Page 98

5020A
Users Manual
6-28
DATa:DSTatistics:RECord:ETIMe? [
Returns the daily statistics recording end time.
Example: DAT:DST:REC:ETIM?
Response: 23,59,59
The
record 1 is assumed. The response is returned in the format,
DATa:DSTatistics:RECord:FIND? (
Returns the daily statistics record number for the given date.
Example: DAT:DST:REC:FIND? 2003,9,13
Response: 25
If no daily statistics were recorded for the given date a 0 is returned.
DATa:DSTatistics:RECord:VALue?
Returns the daily statistics statistical calculation.
Example: DAT:DST:REC:VAL? 1,1,1
Response: 25.559,0.733,23.097,27.203,4.106,43200,17.16,0
The
specifies the channel (1 or 2). The
standard deviation (2), minimum (3), maximum (4), spread (5), maximum rate (7), or
alarm count (8). Temperature values are in the current temperature unit, C of F.
DATa:RECord:CLEar
Clears all recorded measurement data.
Example: DAT:REC:CLE
This command is password protected. To access this command when the
SYST:PASS:REC is enabled, the password must be successfully entered using the
SYST:PASS:CEN
DATa:RECord:FEED:RHUMidity
Returns the state of the humidity recording enable for the specified channel.
Example: DAT:REC:FEED:RHUM1?
Response: 1
The RHUMidity suffix
the humidity recording for the specified channel is enabled. A value of 0 is returned if the
humidity recording for the specified channel is disabled.
DATa:RECord:FEED:RHUMidity
Sets the state of the humidity recording enable for the specified channel.
Example: DAT:REC:FEED:RHUM1 0
The RHUMidity suffix
the humidity recording for the specified channel on (1 or ON) or off (0 or OFF). This
command may be password protected (see Password Commands).