beautypg.com

Fluke 5020A User Manual

Page 108

background image

5020A

Users Manual

6-38

A value of 1 is returned if serial port echo (duplex) is ON. A value of 0 is returned if

serial port echo (duplex) is OFF.

SYSTem:COMMunicate:SERial:FDUPlex

Sets the serial port echo (duplex) setting.
Example: SYST:COMM:SER:FDUP 1
The parameter turns serial port echo (duplex) mode on (1 or ON) or off (0 or

OFF). A baud rade of 57600 suspends echo (full duplex).

SYSTem:COMMunicate:SERial:FEED?

Returns the state of the serial port auto printing.
Example: SYST:COMM:SER:FEED?
Response: 0
A value of 1 is returned if serial port auto printing is enabled. A value of 0 if returned is

serial port auto printing is disabled.

SYSTem:COMMunicate:SERial:FEED

Sets the state of the serial port auto printing.
Example: SYST:COMM:SER:FEED 1
The parameter turns serial port auto printing on (1 or ON) or off (0 or OFF). The

*RST command sets the serial port auto printing to OFF.

SYSTem:COMMunicate:SERial:LINefeed?

Returns the state of the serial port linefeed setting.
Example: SYST:COMM:SER:LIN?
Response: 1
A value of 1 is returned if the serial port linefeed is enabled. A value of 0 is returned if

serial port linefeed is disabled.

SYSTem:COMMunicate:SERial:LINefeed

Sets the state of the serial port linefeed setting.
Example: SYST:COMM:SER:LIN 1
The parameter turns serial port linefeed on (1 or ON) or off (0 or OFF).

SYSTem:COMMunicate:SERial:TIMe? [MIN|MAX|DEF]

Returns the serial port auto printing period in seconds.
Example: SYST:COMM:SER:TIM?
Response: 10
Specifying the MIN, MAX, or DEF parameter returns the minimum, maximum, or

default value respectively.

SYSTem:COMMunicate:SERial:TIMe |MIN|MAX|DEF

Sets to serial port sample period setting in seconds.