Fluke 5020A User Manual
Page 93

Digital Communications Interface
Commands
6
6-23
CALCulate[
Returns the specified statistical calculation for the specified channel.
This is the same as the CALCulate,chn.:PARameter
(see ).
CALCulate[
Returns the name of the specified statistical type for the specified channel.
Example: CALC1:PAR1:AVER4:TYPE?
Response: “T MAX”
The CALCulate suffix,
specifies the statistic: average (1), standard deviation (2), minimum (3), maximum (4),
spread (5), maximum rate (7), or alarm count (8). Temperature values are in the current
temperature unit, C of F.
CALCulate
Returns the rate calculation for the specified channel in degrees C or F per hour for
temperature or % per hour for relative humidity.
Example: CALC1:PAR1:RATE?
Response: -0.74
The CALCulate suffix,
CALCulate[
Returns the rate time span in seconds.
Example: CALC:PAR:RATE:TIME?
Response: 300
Specifying the MIN, MAX, or DEF parameter returns the minimum, maximum, or
default value respectively.
CALCulate[
Sets the rate time span for the specified channel in seconds.
Example: CALC:PAR:RATE:TIME 600
Specifying MIN, MAX, or DEF parameter sets the rate time span to the minimum,
maximum, or default value respectively. This command may be password protected (see
Password Commands).
CALCulate[
Returns the display resolution.
Example: CALC:PAR1:RES?
Response: 3
The PARameter suffix,
humidity resolution (2). The value returned indicates the number of decimal places.