Digital Communications Interface

Commands

6

6-31

Example: SENS:AVER?
Response: 1
A value of 1 is returned if the measurement averaging is enabled. A value of 0 is returned

if the measurement averaging is disabled.

SENSor:AVERage

Sets the measurement average setting.
Example: SENS:AVER 1
The parameter turns the measurement averaging on (1 or ON) or off (0 or OFF).

This command may be password protected (see Password Commands).

SENSor:LOCK?

Returns the sensor lock state for the given channel number.
Example: SENS1:LOCK?
Response: 1
A value of 1 is returned if the sensor lock is enabled. A value of 0 is returned if the sensor

lock is disabled.

SENSor:LOCK

Sets the sensor lock for the specified channel.
Example: SENS1:LOCK 1
The parameter turns the sensor lock on (1 or ON) or off (0 or OFF).
This command is password protected. To access this command when the

SYST:PASS:SENS is enabled, the password must be successfully entered using the

SYST:PASS:CEN command (see Password Commands).

SENSor:IDENtification?

Returns the sensor identification for the specified channel.
Example: SENS1:IDEN?
Response: “LABORATORY”
The SENSor suffix, , specifies the channel (1 or 2). The response to this command

is enclosed in double quotes and indicates the current ID string for the sensor on the

specified channel.

SENSor:IDENentification

Sets the sensor identification for the specified channel.
Example: SENS1:IDEN CHAMBER3
The SENSor suffix, , specifies the channel (1 or 2). The sensor identification field

is set to the value indicated by . The sensor identification can be up to 16 characters

in length and consist of any combination of numbers, letters, underscores, and spaces.
This command may be password protected (see Password Commands)