Fluke 5020A User Manual
Page 90

5020A
Users Manual
6-20
ALARm:TEMPurature
Returns the temperature rate alarm limit for the specified channel in degrees (C or F) per
hour.
Example: ALAR:TEMP1:RATE:LIM?
Response: 5
The TEMPerature suffix,
always in the current temperature units (C or F). Specifying the MIN, MAX, or DEF
parameter returns the minimum, maximum, or default value respectively.
ALARm:TEMPurature
Sets the temperature rate alarm limit for the specified channel in degrees (C or F) per
hour.
Example: ALAR:TEMP1:RATE:LIM 10.00
The TEMPerature suffix,
specifies the value for the temperature rate alarm limit. The
current temperature units (C or F). Specifying MIN, MAX, or DEF parameter sets the
temperature rate alarm limit to the minimum, maximum, or default value respectively.
This command may be password protected (see Password Commands).
ALARm:TEMPurature
Indicates whether a temperature sensor alarm event has occurred for the specified
channel.
Example: ALAR:TEMP1:SENS?
Response: 0
The TEMPerature suffix,
a temperature sensor alarm event has occurred on the specified channel. A value of 0 is
returned if a temperature sensor alarm event has not occurred on the specified channel.
ALARm:TEMPurature
Returns the state of the temperature sensor alarm enable for the specified channel.
Example: ALAR:TEMP1:SENS:ENAB?
Response: 0
The TEMPerature suffix,
the temperature sensor alarm is enabled. A value of 0 is returned if the temperature sensor
alarm is disabled.
ALARm:TEMPurature
Sets the state of the temperature sensor alarm enable for the specified channel.
Example: ALAR:TEMP1:SENS:ENAB 1
The TEMPerature suffix,
turns the temperature sensor alarm on (1 or ON) or off (0 or OFF). This command may be
password protected (see Password Commands).
ALARm:TEMPurature
Indicates whether an upper temperature alarm event has occurred.