Digital Communications Interface

Commands

6

6-15

Response: 0
A value of 1 is returned if a power loss alarm event has occurred. A value of 0 is returned

if a power loss alarm event has not occurred.

ALARm:POWer:ENABle?

Returns the state of the power loss alarm enable.
Example: ALAR:POW:ENAB?
Response: 0
A value of 1 is returned if the power loss alarm is enabled. A value of 0 is returned if the

power loss alarm is disabled.

ALARm:POWer:ENABle

Sets the state of the power loss alarm enable.
Example: ALAR:POW:ENAB 1
The parameter turns the power loss alarm on (1 or ON) or off (0 or OFF). This

command may be password protected (see Password Commands).

ALARm:RHUMidity:LOWer?

Indicates whether a lower humidity alarm event has occurred for the specified channel.
Example: ALAR:RHUM1:LOW?
Response: 0
The RHUMidity suffix, , specifies the channel (1 or 2). A value of 1 is returned if a

lower humidity alarm event has occurred on the specified channel. A value of 0 is

returned if a lower humidity alarm event has not occurred on the specified channel.

ALARm:RHUMidity:LOWer:ENABle?

Returns the state of the lower humidity alarm enable for the specified channel.
Example: ALAR:RHUM1:LOW:ENAB?
Response: 0
The RHUMidity suffix, , specifies the channel (1 or 2). A value of 1 is returned if

the lower humidity alarm is enabled. A value of 0 is returned if the lower humidity alarm

is disabled.

ALARm:RHUMidity:LOWer:ENABle

Sets the state of the lower humidity alarm enable for the specified channel.
Example: ALAR:RHUM1:LOW:ENAB 1
The RHUMidity suffix, , specifies the channel (1 or 2). The parameter

turns the lower humidity alarm enable on (1 or ON) or off (0 or OFF). This command

may be password protected (see Password Commands).

ALARm:RHUMidity:LOWer:LIMit? [MIN|MAX|DEF]

Returns the lower humidity alarm limit for the specified channel in %RH.
Example: ALAR:RHUM1:LOW:LIM?